![National Instruments NI 4350 Скачать руководство пользователя страница 57](http://html1.mh-extra.com/html/national-instruments/ni-4350/ni-4350_user-manual_3462206057.webp)
Appendix A
Specifications
©
National Instruments Corporation
A-7
Open-thermocouple detection .............10 M
Ω
between CH+ and
+2.5 V (software selectable)
Ground-referencing.............................10 M
Ω
between CH– and
ground (software selectable)
Input bias current ................................<500 pA
CMR (DC, 50 Hz, 60 Hz, 400 Hz)
Range
≥
2.5 V ..............................80 dB
Range < 2.5 V ..............................100 dB
NMR (50 Hz, 60 Hz, 400 Hz) .............> 100 dB
Dynamic Characteristics
Bandwidth ..........................................20 Hz
Step response (full-scale step
)
Excitation
Number of channels ............................1
Level ..................................................25
µ
A
Maximum load resistance ...................600 k
Ω
Temperature coefficient ......................±15 ppm/° C
Accuracy
Time (s)
±0.1%
0.3
±0.01%
0.5
±0.0015%
2.4
±0.001%
3
±0.0004%
7