Appendix A
Specifications
©
National Instruments Corporation
A-3
RTD Accuracy (with I
EX
or I
EX0
)
1
RTD Accuracy (with I
EX1
)
2
1
RTD specifications assume that the 25 k
Ω
(25 µA current source) range is used and worst case common-mode voltage for this
range is present. Specifications improve if actual common-mode voltage is less than worst case. Specifications improve for a
1,000
Ω
RTD.
Error (°C)
15 °C–35 °C, 1 Year
0 °C–15 °C,
35 °C–55 °C
Filter Setting
Temperature
Coefficient
(°C/°C)
RTD
°C
10 Hz
50 Hz
60 Hz
Pt
100
Ω
–200
1.00
1.33
1.81
0.01
0
1.14
1.49
2.00
100
1.22
1.58
2.10
300
1.38
1.76
2.32
600
1.66
2.08
2.69
2
RTD specifications assume that the 625
Ω
range (1 mA current source) is used and worst case common-mode voltage for this
range is present. Specifications improve if actual common-mode voltage is less than worst case.
Error (°C)
15 °C–35 °C, 1 Year
0 °C–15 °C,
35 °C–55 °C
Filter Setting
Temperature
Coefficient
(°C/°C)
RTD
°C
10 Hz
50 Hz
60 Hz
Pt 100
Ω
–200
0.05
0.06
0.07
0.01
0
0.12
0.13
0.14
100
0.16
0.17
0.18
300
0.23
0.24
0.26
600
0.36
0.37
0.39