Appendix A
Specifications
©
National Instruments Corporation
A-9
100 MHz System Reference Clock: PXIe_CLK100 and
PXIe_SYNC100
Maximum slot-to-slot skew ................... 100 ps
Accuracy ................................................ ±25 ppm max. (guaranteed over
the operating temperature range)
Maximum jitter ...................................... 3 ps RMS phase-jitter
(10 Hz–12 kHz range)
2 ps RMS phase-jitter
(12 kHz–20 MHz range)
Duty-factor for PXIe_CLK100 .............. 45%–55%
Absolute single-ended voltage swing
(When each line in the differential pair
has 50 W termination to 1.30 V
or Thévenin equivalent) ......................... 400–1000 mV
Note
For other specifications refer to the
PXI-5 PXI Express Hardware Specification
.
External 10 MHz Reference Out (BNC on rear panel of
chassis)
Accuracy ................................................ ±25 ppm max. (guaranteed over
the operating temperature range)
Maximum jitter ...................................... 5 ps RMS phase-jitter
(10 Hz–1 MHz range)
Output amplitude.................................... 1 V
PP
±20% square-wave
into 50
Ω
2 V
PP
unloaded
Output impedance .................................. 50
Ω
±5
Ω