Appendix A
Specifications
A-2
ni.com
Gain = 100
Calibrated
1
................................±25
μ
V max
±10
μ
V typ
With autozero enabled
2
............±10
μ
V max
±5
μ
V typ
Gain error (relative to calibration reference)
Gain = 1 or 100
Calibrated
................................0.074% of reading max
0.02% of reading typ
RTD Measurement Accuracy
Amplifier Characteristics
Input coupling.........................................DC
Input impedance
Normal powered on .........................>1 G
Ω
Input bias current ....................................±
2.8
nA
1
Assumes 1,000 point average, 25
°
C ±10
°
C over one year.
2
Assumes 1,000 point average, ±1
°
C of autozero temperature.
Table A-1.
RTD Measurement Accuracy
Measured
Temperature
°
C
100
Ω
Max
°
C
100
Ω
Typ
°
C
1000
Ω
Max
°
C
1000
Ω
Typ
°
C
–100 to 0
0.60
0.23
1.09
0.46
0 to 25
0.62
0.23
1.11
0.47
25 to 100
0.69
0.25
1.20
0.49
100 to 500
1.11
0.37
1.68
0.65
500 to 1200
2.06
0.65
2.81
1.04
Notes
: The accuracies in this table reflect using the module in4-wire mode. They do not include errors from the RTD
including lead-wire errors when using 2- or 3-wire connection.
The accuracies assume auto-zero is enabled and the environmental conditions are 25
°
C ±10
°
C over a one year period.
These accuracies were computed using a standard RTD with a TCR of 3851.