7.
Unplug the calibration laser USB cable from the USB port.
8.
Remove the laser alignment fixture from the measurement antenna fixture.
After antenna alignment is complete, you can perform system calibration and determine all the
path losses in the system.
Related Information
mmWave OTA Test Positioner SFP
on page 25
Performing System Calibration
Use the RF System Calibration Assistant to perform a full system calibration on all the
mmWave OTA Validation Test Reference Architecture hardware components.
NI recommends performing system calibration in any of the following situations:
•
After any module in the chassis is installed, uninstalled, or moved
•
After inter-module cabling has been changed, reconnected, or repositioned
•
After changing controllers, computers, or reinstalling software
•
After self-calibration is performed on the individual instruments
The following equipment is required for the system calibration procedure:
•
PXIe-5831 Vector Signal Transceiver and mmRH-5582 mmWave Radio Head
•
Rohde & Schwarz NRP50S Power Meter
Note
The Rohde & Schwarz NRP50S Power Meter requires software that is not
included with the mmWave OTA Validation Test Software. Install the power meter
driver software before you complete the following steps.
1.
Connect the power meter USB cable to the PXI chassis USB port.
2.
Open the RF System Calibration Assistant and click
Start Calibration
.
3.
In the
Load Configuration Dialog
dialog box, click the file button and browse to
C:\Users\Public\Public Documents\National Instruments\OTA Test
\System Calibration Files
.
4.
Select
OTA-Test-Cal-Base.cfg
and click
Open
.
Note
The calibration configuration assumes that the name of the PXIe-5831
in Measurement & Automation Explorer (MAX) is
3622
. If you are not using
3622
as the name of the PXIe-5831, either change the name in the
configuration file or change it in MAX.
5.
Select
Use existing calibration configuration
.
The
NI RF System Calibration Assistant Configurator
window launches.
6.
In the
Resources
window, verify that all resources and resource settings are correct for
the path loss characterization as shown in the following figure. Ensure the resource name
for the power meter matches the name used in MAX.
18
|
ni.com
|
mmWave OTA Validation Test Reference Architecture