NI 4071 7½-Digit FlexDMM Calibration Procedure
|
© National Instruments
|
57
3.
Wait two minutes for the thermal EMF to stabilize if the Fluke 5440 cables were not
previously connected in this configuration.
4.
Call the niDMM Reset VI.
Table 9.
Fluke 5440 Cable Connections
Fluke 5440
Cable
Identification
Banana Plug
Connector
(NI 4071)
Banana Plug
Color (Fluke
5440 Cable)
Binding Post
(Fluke 5720A
Calibrator)
First cable
HI
Red
OUTPUT HI
LO
Black
OUTPUT LO
(No connection)
Blue
V GUARD
Second cable
HI SENSE
Red
SENSE HI
LO SENSE
Black
SENSE LO
(No connection)
Blue
V GUARD
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_reset
with the
following parameter:
Instrument_Handle
:
The instrument
handle from
niDMM_init