NI 4070/4072 Calibration Procedure
30
ni.com
Verifying DC Current
To verify the DC current of the NI 4070/4072, complete the following
steps:
1.
Reset the calibrator.
2.
Fasten the connectors on one end of the Fluke 5440 cable to the
NI 4070/4072
HI SENSE
and
LO
banana plug connectors, and connect
the connectors on the other end of the cable to the
HI
and
LO
calibrator
binding posts. Figure 4 shows the correct connections. Table 11 lists
the cable connections.
Figure 4.
Cable Connections for Current
3.
Call
niDMM_reset
to reset the NI 4070/4072 to a known state.
4.
Set the current output on the calibrator to NORM and output 0 A.