NI 4070/4072 Calibration Procedure
Figure 2.
Cable Connections for 4-Wire Resistance
3.
Wait two minutes for the thermal EMF to stabilize if the Fluke 5440
cables were not previously connected in this configuration.
4.
Call
niDMM_reset
.
1
NI 4070/4072
2
Fluke 5700A/5720A Calibrator
3
Fluke 5440 Cables
Table 8.
Fluke 5440 Cable Connections
Fluke 5440
Cable Identification
Banana Plug
Connector
(NI 4070/4072)
Banana Plug Color
(Fluke 5440 Cable)
Binding Post
(Fluke 5700A/5720A
Calibrator)
First cable
HI
Red
OUTPUT HI
LO
Black
OUTPUT LO
(No connection)
Blue
V GUARD
Second cable
HI SENSE
Red
SENSE HI
LO SENSE
Black
SENSE LO
(No connection)
Blue
V GUARD