NI 9144 User Guide and Specifications
30
ni.com
Table 17 contains the default scan list.
NI 9205/9206 Calibration Data
The NI 9205 uses a quadratic formula for conversion from 16-bit raw data
to calibrated data.
The NI 9205 EEPROM provides overall polynominal values a3–a0 along
with gain and offset values for each voltage range, to be applied when
converting 16-bit raw data to calibrated data.
1.
Convert the 32-bit hex values to 64-bit floating point format for use in
the calibration formula.
2.
Select the 32-bit gain value for a particular range.
Table 17.
NI 9205/9206 Scan List Format
Index
Sub
Type
Value
Sub
Value
0x2001
0
ARR:U32
33
—
—
1
32
—
—
2
0x00002A38
18
0x00103238
3
0x00012B38
19
0x00113338
4
0x00022C38
20
0x00123438
5
0x00032D38
21
0x00133538
6
0x00042E38
22
0x00143638
7
0x00052F38
23
0x00153738
8
0x0006283C
24
0x0016303C
9
0x0007293C
25
0x0017313C
10
0x00082A3C
26
0x0018323C
11
0x00092B3C
27
0x0019333C
12
0x000A2C3C
28
0x001A343C
13
0x000B2D3C
29
0x001B353C
14
0x000C2E3C
30
0x001C363C
15
0x000D2F3C
31
0x001D373C
16
0x000E3038
32
0x001E2838
17
0x000F3138
33
0x001F2938