©
National Instruments Corporation
41
NI 9144 User Guide and Specifications
The calibration data is stored in a U32 array, though each Offset field
should be interpreted as a signed value.
The NI 9219 returns calibrated 24-bit (padded to 32-bits) AI data for all
modes and ranges.
To convert raw data into engineering units use the following formula:
y
=
m
*
x
+
b
•
b
= offset based on range of the device (such as, –60 for ± 60 Volts
Voltage Measurement Range)
•
m
= Gain (Full-Range/ (2
24
)): (such as, 120/(2
24
) for ± 60 Volts Voltage
Measurement Range)
14
3-Wire RTD
Pt1000
15
Pt100
16
Quarter-Bridge
350
Ω
17
120
Ω
18
Half-Bridge
±500 mV/V
19
Reserved
—
20
Full-Bridge
±62.5 mV/V
21
±7.8 mV/V
Table 35.
NI 9219 Calibration Data
Coefficient
Representation
LSB Weight
Unsigned
Offset
Signed
Table 34.
NI 9219 Channel Calibration (Continued)
Entry
Number
Mode
Range