7-40
|
ni.com
Chapter 7
Counters
Routing Signals to Counter
n
HW Arm Input
Any of the following signals can be routed to the Counter
n
HW Arm input:
•
RTSI <0..7>
•
PFI <0..15>
•
AI Reference Trigger (ai/ReferenceTrigger)
•
AI Start Trigger (ai/StartTrigger)
•
PXI_STAR
•
PXIe_DSTAR<A,B>
•
Analog Comparison Event
•
Change Detection Event
A counter’s Internal Output can be routed to a different counter’s HW Arm.
Some of these options may not be available in some driver software.
Counter
n
Sample Clock Signal
Use the Counter
n
Sample Clock (Ctr
n
SampleClock) signal to perform sample clocked
acquisitions and generations.
You can specify an internal or external source for Counter
n
Sample Clock. You can also specify
whether the measurement sample begins on the rising edge or falling edge of Counter
n
Sample
Clock.
If the DAQ device receives a Counter
n
Sample Clock when the FIFO is full, it reports an
overflow error to the host software.
Using an Internal Source
To use Counter
n
Sample Clock with an internal source, specify the signal source and the
polarity of the signal. The source can be any of the following signals:
•
DI Sample Clock (di/SampleClock)
•
DO Sample Clock (do/SampleClock)
•
AI Sample Clock (ai/SampleClock)
•
AI Convert Clock (ai/ConvertClock)
•
AO Sample Clock (ao/SampleClock)
•
DI Change Detection output
Several other internal signals can be routed to Counter
n
Sample Clock through internal routes.
Refer to
Device Routing in MAX
in the
NI-DAQmx Help
or the
LabVIEW Help
for more
information.