CB-27T Calibration Procedure

T

meas

is in degrees Celsius and

a

= 1.295361

×

10

-3

b

= 2.343159

×

10

-4

c

= 1.018793

×

10

-7

8.

Compare T

act

to T

meas

If (T

meas

– 0.06

°

C)

T

act

(T

meas

+ 0.06

°

C), you have verified

the performance of the terminal block temperature sensor.

If T

act

< (T

meas

– 0.06

°

C), the terminal block temperature sensor

is nonfunctional. Return the terminal block to NI for repair or
replacement.

If T

act

> (T

meas

+ 0.06

°

C), the terminal block temperature sensor

is nonfunctional. Return the terminal block to NI for repair or
replacement.

You have completed verifying the performance of the temperature sensor
on your CB-27T terminal block.