CB-27T Calibration Procedure
T
meas
is in degrees Celsius and
a
= 1.295361
×
10
-3
b
= 2.343159
×
10
-4
c
= 1.018793
×
10
-7
8.
Compare T
act
to T
meas
•
If (T
meas
– 0.06
°
C)
≤
T
act
≤
(T
meas
+ 0.06
°
C), you have verified
the performance of the terminal block temperature sensor.
•
If T
act
< (T
meas
– 0.06
°
C), the terminal block temperature sensor
is nonfunctional. Return the terminal block to NI for repair or
replacement.
•
If T
act
> (T
meas
+ 0.06
°
C), the terminal block temperature sensor
is nonfunctional. Return the terminal block to NI for repair or
replacement.
You have completed verifying the performance of the temperature sensor