BNC-2080 Board
Chapter 2
BNC-208X Series User Manual
2-10
© National Instruments Corporation
analog (ACH0 through ACH7) BNC connectors are used. The remaining eight analog (ACH8 through
ACH15) BNC connectors are not used. The MIO-16 board must also be configured for differential
input. The jumper W1 position is irrelevant in DIFF mode.
SE
DIFF
J3
ACH0
J9
ACH8
S2
SE
DIFF
J3
ACH0
J9
ACH8
S2
DIFF Mode
(Factory Default Setting)
SE Mode
Figure 2-7. Switch Configurations for Differential Mode or Single-Ended Mode
Note: All of the switches are required to be in the same position; that is, S1 through S4 and S6 through
S9 must all be in either the DIFF position or in the SE position.
The single-ended mode is configured by flipping the switches next to each pair of BNC connectors to
SE. This must be done to all eight switches in order to have 16 single-ended channels. In the single-
ended mode, all 16 BNC connectors are in use. In the single-ended mode, all 16 BNC cases are tied to
the COMMON signal. The COMMON signal can be switched between AIGND and AISENSE
through jumper W1. See Figure 2-8. AISENSE is tied to COMMON for NRSE mode configuration
and AIGND is tied to COMMON for RSE mode configuration.
RSE
NRSE
W1
• •
W1
AISENSE
AIGND
COMMON
• •
Figure 2-8. W1 Ground Reference Jumper
The three configurations of the BNC-2080 board corresponding to the input mode configuration of the
MIO-16 are shown in Figure 2-9 and Figure 2-10.