Appendix A
Specifications for AT-MIO-16E-10 and AT-MIO-16DE-10
©
National Instruments Corporation
A-19
Calibration
Recommended warm-up time ................ 15 min
Calibration interval ................................ 1 year
External calibration reference ................ >6 and <10V
Onboard calibration reference
Level ............................................... 5.000 V (±3.5 mV) (over full
operating temperature, actual
value stored in EEPROM)
Temperature coefficient .................. ±5 ppm/°C max
Long-term stability ......................... ±15 ppm/
Bus Interface
Type ....................................................... Slave
Power Requirement
+5 VDC (±5%)....................................... 0.7 A
Power available at I/O connector ........... +4.65 VDC to +5.25 VDC
at 1 A
Physical
Dimensions
(not including connectors) ..................... 17.45 by 10.56 cm
(6.87 by 4.16 in.)
I/O connector
AT-MIO-16E-10 ............................. 68-pin male SCSI-II type
AT-MIO-16DE-10 .......................... 100-pin female 0.050 D-type
Maximum Working Voltage
Maximum working voltage refers to the signal voltage plus the
common-mode voltage.
Channel-to-earth..................................... 42 V, Installation Category II
Channel-to-channel ................................ 42 V, Installation Category II
1,000 h