Caution
Transient protection shall be provided that is
set at a level not exceeding 140% of the peak rated
voltage value of 85 V at the supply terminals to the
equipment.
Caution
The system shall only be used in an area of
not more than Pollution Degree 2, as defined in
IEC/EN 60664-1.
Caution
The system shall be mounted in an
ATEX/IECEx-certified enclosure with a minimum
ingress protection rating of at least IP54 as defined in
IEC/EN 60079-15.
Caution
The enclosure must have a door or cover
accessible only by the use of a tool.
Electromagnetic Compatibility Guidelines
This product was tested and complies with the regulatory
requirements and limits for electromagnetic compatibility (EMC)
stated in the product specifications. These requirements and
limits provide reasonable protection against harmful interference
when the product is operated in the intended operational
electromagnetic environment.
NI 9262 Getting Started Guide
|
© National Instruments
|
7