NI 9230/9232 Calibration Procedure

|

© National Instruments

|

5

b.

Configure the AI voltage task timing according to Table 4.

c.

Start the task. 

d.

Average the samples acquired for each channel.

e.

Stop and Clear the task.

6.

Compare the average to the limits in Table 5.

Note

The test limits listed in Table 5 are derived using the values in Table 7.

Note

The uncertainty of the calibration standard is not included in these limits.

7.

Set the calibrator to Standby mode (STBY).

8.

Repeat steps through 7 for each test point in Table 5.

9.

Disconnect the calibrator from the device.

Adjustment

The following performance adjustment procedure describe

s

 the sequence of operation required 

to adjust the NI 9230/9232.

Accuracy Adjustment

Complete the following procedure to adjust the accuracy performance of the NI 9230/9232.
1.

Set the calibrator to Standby mode (STBY).

2.

Connect the NI 9230/9232 to the calibrator as shown in Figure 1.

3.

Adjust the NI 9230/9232.
a.

Initialize a calibration session on the NI 9230/9232. The default password is 

NI

.

b.

Input the external temperature in degrees Celsius.

Table 4.  

NI 9230/9232 Timing Configuration

Sample Mode

Samples Per Channel

Rate kS/s

NI 9230 Finite Samples

1280

12.8

NI 9232 Finite Samples

10240

102.4

Table 5.  

NI 9230/9232 Verification Test Limits for Accuracy

Range

Test Point

1-Year Limits

Min (V)

Max (V)

Location

Value (V)

Lower Limit (V)

Upper Limit (V)

-30

30

Max

28.5

28.46853

28.53147

Mid

0.00

-0.007245

0.007245

Min

-28.5

-28.53147

-28.46853

Содержание 9232

Страница 1: ...bration Conditions 6 Worldwide Support and Services 7 Software Calibrating the NI 9230 9232 requires the installation of NI DAQmx on the calibration system Table 1 shows the necessary versions for eac...

Страница 2: ...specifications and calibration interval NI 9232 Getting Started Guide NI 9232 specific information NI 9232 Datasheet NI 9232 specifications and calibration interval NI DAQmx Readme Operating system a...

Страница 3: ...ambient temperature Keep relative humidity below 80 Allow a warm up time of at least 10 minutes to ensure that the NI 9230 9232 measurement circuitry is at a stable operating temperature Initial Setup...

Страница 4: ...lete the following procedure to determine the As Found status of the NI 9230 9232 1 Set the calibrator to Standby mode STBY 2 Connect the NI 9230 9232 to the calibrator as shown in Figure 1 Figure 1 A...

Страница 5: ...s the sequence of operation required to adjust the NI 9230 9232 Accuracy Adjustment Complete the following procedure to adjust the accuracy performance of the NI 9230 9232 1 Set the calibrator to Stan...

Страница 6: ...rform an adjustment you can update the calibration date and onboard calibration temperature without making any adjustments by initializing an external calibration setting the C Series calibration temp...

Страница 7: ...e National Instruments website is your complete resource for technical support At ni com support you have access to everything from troubleshooting and application development self help resources to e...

Отзывы: