NI 9216/9226 Calibration Procedure

|

© National Instruments

|

5

Note

The limits are only based on the NI 9216/9226 accuracy. The uncertainty of 

calibration standard is not included.

Note

The test limits listed in Table 2 are derived using the values in Table 6.

5.

Set the calibrator to Operate mode (OPR).

6.

Acquire and average samples from one channel.
a.

Create and configure a 4-wire AI resistance task on the NI 9216/9226 according to 
Table 3.

b.

Configure the ADC Timing Mode to High Resolution.

Note

Use appropriate channel number in place of “x” in “AIx.” For example, use 

AI3 for channel 3.

c.

Configure the task timing according to Table 4.

d.

Start the task.

e.

Read the samples from the NI 9216/9226.

f.

Average the samples.

g.

Stop and clear the task.

7.

Compare the per-channel averages to the limits in Table 2.

8.

Set the calibrator to Standby mode (STBY).

9.

Repeat steps 3 through 8 for each test point listed in Table 2.

10. Disconnect the NI 9216/9226 channel from the calibrator.
11. Repeat steps 2 through 10 for any other desired channel of NI 9216/9226. 

Table 3.  

NI 9216/9226 Voltage Configuration

Module

Physical Channel

Range

Minimum (

Ω

)

Maximum (

Ω

)

NI 9216

AIx

0

400

NI 9226

AIx

0

4000

Table 4.  

NI 9216/9226 Task Timing Configuration

Sample Mode

Samples Per Channel

Rate (S/s)

Finite Samples

10

5

Содержание 9216

Страница 1: ... 2 Test Equipment 3 Test Conditions 3 Initial Setup 3 Verification 4 Accuracy Verification 4 Adjustments 6 Accuracy Adjustment 6 EEPROM Update 7 Reverification 7 Accuracy Under Calibration Conditions 7 Worldwide Support and Services 8 Software Calibrating the NI 9216 9226 requires the installation of NI DAQmx 14 5 or later on the calibration system You can download NI DAQmx from ni com downloads N...

Страница 2: ...t NI 9216 circuitry and specifications NI 9226 Getting Started Guide NI 9226 connection information NI 9226 Datasheet NI 9226 circuitry and specifications NI DAQmx Readme Operating system and application software support in NI DAQmx LabVIEW Help LabVIEW programming concepts and reference information about NI DAQmx VIs and functions NI DAQmx C Reference Help Reference information for NI DAQmx C fun...

Страница 3: ...midity below 80 Allow a warm up time of at least 10 minutes to ensure that the NI 9216 9226 measurement circuitry is at a stable operating temperature Initial Setup Complete the following steps to set up the NI 9216 9226 1 Install NI DAQmx 2 Make sure the cDAQ 9178 power source is not connected to the chassis 3 Connect the cDAQ 9178 chassis to the system safety ground a Attach a ring lug to a 14 A...

Страница 4: ...ing procedure to determine the As Found status of the NI 9216 9226 1 Set the calibrator to Standby mode STBY 2 Connect the desired channel of the NI 9216 9226 to the calibrator as shown in Figure 1 Connect only one channel at a time Figure 1 4 Wire Resistance Connections 3 Set the calibrator as a resistance source to a Test Point value indicated in Table 2 4 Set calibrator COMP to 4 wire Table 2 N...

Страница 5: ... in AIx For example use AI3 for channel 3 c Configure the task timing according to Table 4 d Start the task e Read the samples from the NI 9216 9226 f Average the samples g Stop and clear the task 7 Compare the per channel averages to the limits in Table 2 8 Set the calibrator to Standby mode STBY 9 Repeat steps 3 through 8 for each test point listed in Table 2 10 Disconnect the NI 9216 9226 chann...

Страница 6: ...tain an array of recommended calibration resistance values d Set the calibrator to a reference value determined by the array of recommended calibration voltages e Set calibrator COMP to 4 wire f Set the calibrator to Operate mode OPR g Call and configure the NI 9216 9226 adjustment function according to Table 5 Note Use appropriate channel number in place of x in AIx For example use AI3 for channe...

Страница 7: ...justment verify that you have met the Test Conditions before returning your device to NI Refer to Worldwide Support and Services for assistance in returning the device to NI Accuracy Under Calibration Conditions The values in the following table are based on calibrated scaling coefficients which are stored in the onboard EEPROM The following accuracy table is valid for calibration under the follow...

Страница 8: ...ained in this manual was developed at private expense and is subject to the applicable limited rights and restricted data rights as set forth in FAR 52 227 14 DFAR 252 227 7014 and DFAR 252 227 7015 Worldwide Support and Services The NI website is your complete resource for technical support At ni com support you have access to everything from troubleshooting and application development self help ...

Отзывы: