NI 9215E Operating Instructions and Specifications
Connecting Single-Ended Voltage Signals to the
NI 9215E
To connect single-ended voltage signals to the NI 9215E with
screw terminal, you must also connect the ground signal to the
COM terminal to keep the common-mode voltage in the specified
range, as shown in Figure 6. For more information about the
common-mode voltage range, refer to the
section.
Figure 6.
Connecting a Single-Ended Voltage Signal to the
NI 9215E with Screw Terminal
+
–
AI+
AI–
COM
Voltage
Source
NI 9215E with
Screw Terminal