NI 9205/9206 Calibration Procedure
14
ni.com
Test Point
The
Test Point
is the voltage value that is input or output for verification
purposes. This value is broken down into two columns—
Location
and
Value
.
Location
refers to where the test value fits within the test range.
Value
refers to the voltage value to be verified.
Max
refers to maximum
value,
Min
refers to minimum value, and
Mid
refers to mid-scale.
2-Year Limits
The
2-Year Limits
column contains the
Upper Limits
and
Lower Limits
for
the test point value. That is, when the device is within its 2-year calibration
interval, the test point value should fall between these upper and lower limit
values.
Table 4.
NI 9205/9206 Analog Input Values
Range (V)
Test Point
2-Year Limits
Minimum
Maximum
Location
Value (V)
Lower Limit (V)
Upper Limit (V)
–10
10
Max
9.980000
9.976836
9.983164
–10
10
Mid
0.000000
–0.001407
0.001407
–10
10
Min
–9.980000
–9.983164
–9.976836
–5
5
Max
4.990000
4.988303
4.991697
–5
5
Mid
0.000000
–0.000718
0.000718
–5
5
Min
–4.990000
–4.991697
–4.988303
–1
1
Max
0.998000
0.997617
0.998383
–1
1
Mid
0.000000
–0.000168
0.000168
–1
1
Min
–0.998000
–0.998383
–0.997617
–0.200
0.200
Max
0.199600
0.199489
0.199711
–0.200
0.200
Mid
0.000000
–0.000056
0.000056
–0.200
0.200
Min
–0.199600
–0.199711
–0.199489