Chapter 2
Configuration and Installation
©
National Instruments Corporation
2-3
The jumpers are used as follows:
•
Fixed jumpers
–
On Revision A and B SCXI-1120 modules, jumper W42 is unused
and should not be connected.
–
Jumper W45 is reserved and should not be reconfigured.
–
On Revision A and B SCXI-1120 modules, jumper W44 carries
the SLOT0SEL* signal from the rear signal connector, after
buffering, to the SCXIbus INTR* line and should be left in the
SCXI-1120 factory-default position (position 1). On all
SCXI-1120D modules and on Revision C and later SCXI-1120
modules, jumper W44 does not exist.
Note
The revision letter is located on the National Instruments label located on the side of
each SCXI-1120/D module. The revision letter is found in the part number and is the letter
found in that part number.
•
User-configurable jumpers
–
Jumper W43 carries the SCXIbus MISO line, after buffering, to
the SERDATOUT signal on the rear signal connector.
–
On all SCXI-1120D and on Revision C and later SCXI-1120
modules, jumper 42 connects a pullup resistor to the
SERDATOUT signal on the rear signal connector.
–
Jumper W46 configures the guard, the analog output ground, and
enables the NRSE mode.
–
Jumpers W1 through W8 configure the first-stage gain of channels
0 through 7, respectively.
–
Jumpers W9 through W16 configure the second-stage gain of
channels 0 through 7, respectively.
–
Jumpers W17 through W24 configure the first-stage filtering of
input channels 0 through 7, respectively.
–
Jumpers W25 through W40 configure the second-stage filtering of
input channels 0 through 7, respectively.
–
Jumper W41 directly connects the temperature sensor to the rear
signal connector. To do so, set jumper W46 in the AB-R2 or
AB-R0 position first.