NI 6110/6111/6115/6120 Calibration Procedure
8
ni.com
Verifying the Performance of the NI 6110/6111/6115/6120
This section describes the verification procedures for calibrating the
NI 6110/6111/6115/6120. You must verify analog input, analog output,
and the counter performance.
Verifying Analog Input
To verify the performance of the analog input of the device, complete the
following steps:
1.
Read the
Equipment and Other Test Requirements
section.
Note
Calibrator connections depend on the channel you are verifying. The verification
process requires you to individually test each channel.
2.
Use the following guide to determine the connections between the
calibrator and the device pins for analog input verification.
a.
Starting with channel 0, connect the positive output of the
calibrator to the analog input specified in Table 1.
b.
Connect the negative output of the calibrator to the analog input
specified in Table 2.
Table 1.
Positive Output Connections
Device Channel
Analog Input
0
68
1
33
2
65
3
30
Table 2.
Negative Output Connections
Device Channel
Analog Input
0
34
1
66
2
31
3
63