Clocking

The NI 5761 clocks control the sample rate and other timing functions on the device. Table 3 contains 
information about the possible NI 5761 clock resources.

Using Your NI 5761R with a LabVIEW FPGA Example VI

The NI FlexRIO Adapter Module Support software includes a variety of example projects to help get 
you started creating your LabVIEW FPGA program. This section explains how to use an existing 
LabVIEW FPGA example project to generate and acquire samples with the NI 5761R. This example 
requires at least one SMA cable for connecting signals to your NI 5761R.

Note

Examples available for your device are dependent on the device-specific minimum software 

requirements. For more information about software requirements for your device, visit 

ni.com/

info

 and enter 

rdsoftwareversion

 as the Info Code to determine which minimum software 

versions you need for your device.

Each NI 5761R example project includes the following components:

A LabVIEW FPGA VI that can be compiled and run embedded in FPGA hardware

A VI that runs in LabVIEW for Windows and interacts with the LabVIEW FPGA VI

Note

In NI application software, NI FlexRIO adapter modules are referred to as 

IO Modules

.

Complete the following steps to run an example that acquires a waveform on CH 0 of the NI 5761.

1.

Connect one end of an SMA cable to CH 0 on the front panel of the NI 5761 and the other end of 
the cable to your device under test (DUT).

2.

Launch LabVIEW.

3.

In the 

Getting Started

 window, click 

Find Examples

 to display the NI Example Finder.

4.

In the 

NI Example Finder

 window, select 

Hardware Input and Output»FlexRIO»IO Modules»

NI 5761

.

5.

Select 

NI 5761 - Getting Started.lvproj

.

6.

In the 

Project Explorer

 window, open 

NI 5761 - Getting Started (Host).vi

 under 

My Computer

The host VI opens. This VI uses the NI 7952R as the FPGA target by default. If you are using an 
NI FlexRIO FPGA module other than the NI 7952R, complete the following steps to change to an 
FPGA VI that supports your target.

a.

Select 

Window»Show Block Diagram 

to open the VI block diagram.

b.

On the block diagram, right-click the Open FPGA VI Reference (PXI-7952R) function and 
select 

Configure Open FPGA VI Reference

.

Table 3.  

NI 5761 Clock Sources

Clock

Frequency

Source Options

Sample Clock

175 MHz to 250 MHz

Internal VCO locked to a Reference clock

External through the CLK IN connector

External through IoModSyncClock

Reference Clock

10 MHz

Internal

External through the CLK IN connector

External through IoModSyncClock

Содержание 5761R

Страница 1: ...ts Front Panel and Connector Pinouts 2 Block Diagram 4 NI 5761 Component Level Intellectual Property CLIP 5 Cables 6 Clocking 7 Using Your NI 5761R with a LabVIEW FPGA Example VI 7 Creating a LabVIEW Project and Run a VI on an FPGA Target 9 How to Use Your NI FlexRIO Documentation Set 12 Specifications 13 Where to Go for Support 24 Note Before configuring your NI 5761R you must install the appropr...

Страница 2: ... before powering down the module and only connect signals after the adapter module has been powered on by the NI FlexRIO FPGA module Table 1 NI 5761 Front Panel Connectors Device Front Panel Connector Signal Description AUX I O Refer to Table 2 for the signal names and descriptions D 0 3 LEDs for custom configuration CLK IN Provides the NI 5761 with an external Sample clock or Reference clock TRIG...

Страница 3: ...ssis NI is not liable for any damage resulting from such signal connections For the maximum input and output ratings for each signal refer to the Specifications section of this document Table 2 NI 5761 AUX I O Connector Pin Assignments Micro D Connector Pin Signal Signal Description 1 AUXIO0 General purpose digital input or output channels 2 AUXIO1 3 AUXIO2 4 AUXIO3 5 AUXIO4 6 AUXIO5 7 AUXIO6 8 AU...

Страница 4: ...Initialization Done Reinitialize Configuration Error Sample Clock Select Sample Clock Commit SPI Idle Analog FE IO Module Clock 0 n 1 Single Sample CLIP n 2 Multiple Sample CLIP Multiple Sample CLIP Single Sample CLIP AI 0 AI 1 AI 2 AI 3 AI 0 Data N AI 1 Data N AI 2 Data N AI 3 Data N AI 0 Data N 1 AI 1 Data N 1 AI 2 Data N 1 AI 3 Data N 1 Data Clock Data Sample Clock Clock Buffer Internal Referen...

Страница 5: ... IP integration functionality of the user defined CLIP but also allows the CLIP to communicate directly with circuitry external to the FPGA Adapter module socketed CLIP allows your IP to communicate directly with both the FPGA VI and the external adapter module connector interface Figure 3 shows the relationship between an FPGA VI and CLIP Figure 3 CLIP Relationship Adapter Module CLIP Socket LabV...

Страница 6: ...e CLIP Generates one sample per clock cycle at a default sample rate of 250 MHz You can set a lower sample rate by using an external Sample clock This CLIP provides access to four analog input channels eight PFI lines and an input clock selector that can be configured to use one of the following settings Internal Sample clock Internal Sample clock locked to an external Reference clock through the ...

Страница 7: ...VI Note In NI application software NI FlexRIO adapter modules are referred to as IO Modules Complete the following steps to run an example that acquires a waveform on CH 0 of the NI 5761 1 Connect one end of an SMA cable to CH 0 on the front panel of the NI 5761 and the other end of the cable to your device under test DUT 2 Launch LabVIEW 3 In the Getting Started window click Find Examples to disp...

Страница 8: ... Channel box 9 Set the Trigger Level V and the Record Size controls to the desired value 10 In the Trigger Type box you can select either Software Trigger or Data Edge If you select Software Trigger the VI acquires data every time you click the Software Trigger button on the front panel of the VI If you select Data Edge the VI acquires data every time an edge occurs 11 Click the Run button to run ...

Страница 9: ...125 MHz in the Compile for single frequency control Click OK 8 Right click the FPGA target and select New FPGA Base Clock again 9 In the Resource pull down menu select 200 MHz Clock Click OK 10 Right click IO Module and select Properties In the General category you can see the available CLIP for the NI 5761 in the Component Level IP pane If the category information is dimmed select the Enable IO M...

Страница 10: ... right click My Computer and select New VI A blank VI opens Select Window Show Block Diagram to open the VI block diagram 2 Add the Open FPGA VI Reference function located on the FPGA Interface palette to the block diagram 3 Right click the Open FPGA VI Reference function labeled No Target and select Configure Open FPGA VI Reference 4 In the Configure Open FPGA VI Reference dialog box select the V...

Страница 11: ...I Reference function located on the FPGA Interface palette to the right of the While Loop on the block diagram 17 Wire the FPGA VI Reference Out indicator of the Read Write Control function to the FPGA VI Reference In control of the Close FPGA VI Reference function 18 Wire the error out indicator of the Read Write Control function to the error in control of the Close FPGA VI Reference function You...

Страница 12: ...Help Embedded in LabVIEW Help Contains information about the basic functionality of LabVIEW FPGA Module NI FlexRIO Help Embedded in LabVIEW FPGA Module Help Contains FPGA module adapter module and CLIP configuration information LabVIEW Examples Available in LabVIEW Example Finder Contains examples of how to run FPGA VIs and Host VIs on your device Other Useful Information on ni com ni com ipnet Co...

Страница 13: ...otherwise noted All graphs illustrate the performance of a representative module Typical values describe useful product performance that are not covered by warranty Typical values cover the expected performance of units over ambient temperature ranges of 23 5 C with a 90 confidence level based on measurements taken during development or production Analog Input AI CH 0 through AI CH 3 General Chara...

Страница 14: ... dB 100 1 MHz 90 dB 501 MHz 80 dB AC Coupled Measurements Figure 8 AC Coupled Bandwidth Passband Table 5 AC Coupled Spectral Performance Measurement 20 17 MHz 70 17 MHz 123 17 MHz SNR 72 5 dB 71 4 dB 70 5 dB SINAD 72 3 dB 71 2 dB 70 3 dB SFDR 88 dB 84 dB 80 dB These measurements were extrapolated from a 4 dBFS plot 1M 80k 15 14 13 12 11 10 9 8 7 6 5 4 3 2 1 0 1 2 3 4 5 Gain dB Frequency Hz 100k 10...

Страница 15: ...MS Average Figure 10 AC Coupled Spectral Measurements 20 1 MHz 1 dBFS 8 192 Point FFT 10 RMS Average 20 0 115 110 105 100 95 90 85 80 75 70 65 60 55 50 45 40 35 30 25 20 15 10 5 0 Amplitude dBFS Frequency MHz 40 60 80 100 120 10 0 115 110 105 100 95 90 85 80 75 70 65 60 55 50 45 40 35 30 25 20 0 Amplitude dBFS Frequency MHz 15 10 5 20 30 40 50 60 70 90 100 110 120 80 ...

Страница 16: ...C to 250 MHz Table 6 lists the DC coupled spectral performance measurements All values are measured with a 250 MHz external Sample clock Channel to channel isolation 1 MHz 90 dB 100 1 MHz 80 dB 501 MHz 70 dB Table 6 DC Coupled Spectral Performance Measurement 20 1 MHz 70 1 MHz 122 1 MHz SNR 65 7 dB 64 3 dB 63 dB SINAD 65 2 dB 61 8 dB 56 6 dB SFDR 76 dB 65 dB 58 dB 0 115 110 105 100 95 90 85 80 75 ...

Страница 17: ... Ω you can add series resistance close to the source to properly bias the NI 5761 input terminals You can also use the NI 5761 input channel bias DACs to remove DC offset present in the system For more information about programming the bias DACs refer to the NI 5761 CLIP topics in the NI FlexRIO Help DC Coupled Measurements Figure 12 DC Coupled Bandwidth Passband 10 0 15 14 13 12 11 10 9 8 7 6 5 4...

Страница 18: ...Figure 14 DC Coupled Spectral Measurements 20 1 MHz 1 dBFS 8 192 Point FFT 10 RMS Average 0 115 110 105 100 95 90 85 80 75 70 65 60 55 50 45 40 35 30 25 20 0 Amplitude dBFS Frequency MHz 15 10 5 25 50 75 100 125 0 115 110 105 100 95 90 85 80 75 70 65 60 55 50 45 40 35 30 25 20 0 Amplitude dBFS Frequency MHz 15 10 5 10 20 30 40 50 60 70 80 90 100 110 120 ...

Страница 19: ...nt FFT 10 RMS Average Analog Input Phase Noise Figure 16 Analog Input Phase Noise 0 115 110 105 100 95 90 85 80 75 70 65 60 55 50 45 40 35 30 25 20 0 Amplitude dBFS Frequency MHz 15 10 5 20 40 60 80 100 120 1k 0 160 0 150 0 140 0 130 0 120 0 Phase Noise dBc Hz Frequency Offset from 10 MHz Carrier Hz 110 0 100 0 90 0 80 0 70 0 10k 100k 500k ...

Страница 20: ... Characteristics Number of channels 1 single ended Connector SMA Input impedance 50 Ω Input coupling AC External Sample Clock Input voltage range 0 63 Vpk pk to 2 5 Vpk pk Input frequency range 175 MHz to 250 MHz Absolute maximum input 10 V DC 3 1 Vpk pk AC External Reference Clock Input voltage range 1 4 Vpk pk to 4 4 Vpk pk Input frequency range 10 MHz Absolute maximum input 10 V DC 5 Vpk pk AC ...

Страница 21: ...Zout 50 Ω Iout 2 mA Maximum toggle frequency 500 kHz Absolute maximum input 0 5 V to 7 V EEPROM Map Caution Only write to User Space Writing to any other offset may cause the NI 5761 to stop functioning Voltage Level Minimum Maximum VIL 0 V 0 7 V VIH 1 7 V 5 5 V Voltage Level Minimum Maximum VIL 0 V 0 7 V VIH 1 7 V 5 5 V VOL 0 V 0 4 V VOH 1 9 V 2 5 V Byte Address Size Bytes Field Name 0x0 2 Vendor...

Страница 22: ...e with IEC 60068 2 1 and IEC 60068 2 2 Relative humidity range 5 to 95 noncondensing tested in accordance with IEC 60068 2 56 Note Clean the device with a soft non metallic brush Make sure that the device is completely dry and free from contaminants before returning it to service Safety This product meets the requirements of the following standards of safety for electrical equipment for measuremen...

Страница 23: ...ber or product line and click the appropriate link in the Certification column Environmental Management National Instruments is committed to designing and manufacturing products in an environmentally responsible manner NI recognizes that eliminating certain hazardous substances from our products is beneficial not only to the environment but also to NI customers For additional environmental informa...

Страница 24: ...orth Mopac Expressway Austin Texas 78759 3504 National Instruments also has offices located around the world to help address your support needs For telephone support in the United States create your service request at ni com support and follow the calling instructions or dial 512 795 8248 For telephone support outside the United States contact your local branch office Australia 1800 300 800 Austri...

Отзывы: