NI 4065 Calibration Procedure
4-Wire Resistance
2-Wire Resistance
Table 22.
4-Wire Resistance Verification Tolerances
Calibrator Resistance
Range (
)
Limits, 1-Year (±)
Limits, 24-hour (±)
ppm of Reading
*
ppm of Range
ppm of Reading
*
ppm of Range
0
100
110
40
30
30
10
100
100
100
0
1 k
110
20
20
8
100
1 k
1 k
1 k
0
10 k
110
20
20
8
1 k
10 k
10 k
10 k
0
100 k
110
20
20
8
10 k
100 k
100 k
100 k
0
1 M
125
24
20
12
100 k
1 M
1 M
1 M
*
ppm of Reading applied to displayed discrete resistance value of calibrator.
Table 23.
2-Wire Resistance Verification Tolerances
Calibrator Resistance
Range (
)
Limits, 1-Year (±)
Limits, 24-hour (±)
ppm of Reading
*
ppm of Range
ppm of Reading
*
ppm of Range
10
100
110
2040
30
2030
100
100
100
1 k
110
220
20
208
1 k
1 k
1 k
10 k
110
40
20
28
10 k
10 k