National Instruments Corporation
77
NI 4065 Calibration Procedure
Adjusting DC Current
To adjust DC current for the NI 4065, complete the following steps:
1.
If you are using an NI PXI/PCI/PCIe-4065, fasten the connectors on
one end of the Fluke 5440 cable to the HI SENSE and LO banana plug
connectors, and fasten the connectors on the other end of the cable to
the HI and LO calibrator binding posts, respectively. Figure 7 shows
the correct connections. Table 12 lists the cable connections.
If you are using an NI USB-4065, fasten the connectors on one end of
the Fluke 5440 cable to the I and LO banana plug connectors, and
fasten the connectors on the other end of the cable to the HI and LO
calibrator binding posts, respectively. Figure 8 shows the correct
connections. Table 12 lists the cable connections.
2.
Wait two minutes for the thermal EMF to stabilize.
3.
Call the niDMM Reset VI to reset the NI 4065 to a known state.
4.
Call the niDMM Config Measurement VI. The niDMM Config
Measurement VI is polymorphic. To change the selected instance of
this polymorphic VI, right-click the VI and choose
Select Type
followed by
Resolution in Digits
.
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_reset
with the following
parameter:
Instrument_Handle
:
Cal Session
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_ConfigureMeasurement
Digits
with the following parameters:
Instrument_Handle
:
Cal Session
Resolution_Digits
:
6.5
Measurement_Function
:
NIDMM_VAL_DC_CURRENT
Range
:
1.0