National Instruments Corporation
45
NI 4065 Calibration Procedure
Figure 8.
USB Cable Connections for Current
3.
Wait two minutes for the thermal EMF to stabilize.
4.
Call the niDMM Reset VI to reset the NI 4065 to a known state.
1
NI USB-4065
2
Fluke 5700A/5720A Calibrator
3
Fluke 5440 Cable
Table 12.
Fluke 5440 Cable Connections
Banana Plug Connector
(NI 4065)
Banana Plug Color
(Fluke 5440 Cable)
Binding Post
(Fluke 5700A/5720A Calibrator)
HI SENSE
(NI PXI/PCI/PCIe-4065 only)
Red
OUTPUT HI
LO
Black
OUTPUT LO
I (NI USB-4065 only)
Red
OUTPUT HI
(No connection)
Blue
V GUARD
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_reset
with the following
parameter:
Instrument_Handle
: The instrument
handle from
niDMM_init
AMP
S
3
A
MAX
INPUT
V
Ω
3
00 V
MAX
S
EN
S
E
Ω
4 WIRE
HI
HI
LO
LO
HI
HI
CAT II
CAT II
CAT II
6½-Digit DMM
6½-Digit DMM
6½-Digit DMM
NI USB-4065
NI USB-4065
NI USB-4065
NI USB-4065
3
00 V
MAX
3
00 V
MAX
HI
LO
HI
LO
HI
AUX
CURRENT
GUARD
GROUND
1
2
3