National Instruments Corporation
43
NI 4065 Calibration Procedure
c.
Call the niDMM Read VI.
d.
Verify that the measurement for 0
falls between the tolerances
listed in Table 24.
10. Repeat step 9 for each of the remaining iterations shown in Table 11.
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_read
with the following
parameters:
Instrument_Handle
: The instrument
handle from
niDMM_init
Reading
: Verify that the measurement
for 0
falls between the tolerances listed
in Table 24
Maximum_Time
: –
1
Table 11.
2-Wire Resistance Settings
Iteration
Range
1
100 M
2
10 M
3
1 M
4
100 k
5
10 k
6
1 k
7
100