NASONEB 7070 Скачать руководство пользователя страница 6

6

CLEANING AND STORAGE

After each use, clean the NasoNeb Sinus Therapy System.  

1. Unplug the Compressor

2. Disconnect the Cup Assembly from the Supply Tubing

3.  Disassemble the Cup Assembly and wash the Cup and Insert in warm soapy water

4. Wipe the Supply Tubing and Compressor with a damp cloth

5. Set components aside to dry between uses

6. You may rinse the Cup and Insert with hydrogen peroxide or isopropyl alcohol 

Replace Supply Tubing and Cup Assembly with a Model 7080 NasoNeb Sinus Therapy 

System Supply Kit every 6-12 months or every 730 cleaning cycles. Store the NasoNeb 

Sinus Therapy System in a dry place.

Caution

 – Avoid the use of harsh detergents or cleaning agents. These may damage 

the plastic components, cause fogging or other undesirable effects. Do not wash in 

the dishwasher.

Caution

 – Do not allow water to enter the Supply Tubing.

Caution

 – If water enters or condensation forms in the Supply Tubing, connect 

one end of the Supply Tubing to the Compressor and run the Compressor until the 

condensation evaporates.

REPLACE THE FILTER

The Compressor has a filter installed in the Filter Holder. There are additional filters 

included in the system. Check the filter every 6 months or 720 minutes of use. If it is 

dirty, replace the filter. Additional filters can be acquired from NasoNeb, Inc. or where 

you purchased the NasoNeb System. 

1. Pull the Filter Holder out of the compressor (the Filter Holder pulls straight out)

2. Remove the filter from the Filter Holder

3.  If the filter is discolored, replace with a new filter, otherwise, put the filter back 

into the Filter Holder

4. Insert the Filter Holder back into the compressor before using

Содержание 7070

Страница 1: ...or Use NasoNeb Inc 153 Industrial Blvd Plattsburgh NY 12901 USA Telephone 1 866 960 9833 www nasoneb com MODEL 7070 NasoNeb Sinus Therapy System MODEL 7080 NasoNeb Sinus Therapy System Supply Kit European English EU ...

Страница 2: ...nalsobeused tocleanthenasalpasageswithourproprietaryNasoNeb MoisturizingSolution Ifyouhaveanyquestionsorconcerns pleasecontacttheNasoNeb Inc Customer ExperienceTeamtoll freeintheUSat1 866 960 9833orsendusanemailto hello nasoneb com Acustomerexperienceagentwillbehappytohelpyou Caution Readallwarningsandinstructionsbeforeuse Caution Donotusewithmedicationsthatareincompatiblewith polycarbonate Cautio...

Страница 3: ...sintendeduse asdescribedinthismanual UseonlywithNasoNeb brandaccessories Neverdroporinsertanyobjectintoanyopeningorhose Donotuseoutdoors ThisNasoNebSinusTherapySystemisforindooruseonly 3 Power Supply NasoNeb Compressor On Off Switch Tubing Holder Power Connection Filter Holder Outlet NasoNeb Cup Assembly NasoNeb Supply Tubing NasoNeb Filters ...

Страница 4: ...ConnectoneendoftheSupplyTubingto theoutletoftheNasoNebCompressor 4 ConnectthelooseendoftheSupply TubingtotheairinletoftheCup 5 PlacetheInsertintotheCupto formtheCupAssembly 6 TheInsertandCupfittogetherin onlyoneorientation IftheInsert doesnotseat rotatetheInsert180degrees 7 PlacetheSupplyTubinginthetubingholder toholdtheCupAssemblyupright 8 PlugthepowersupplyintotheCompressor 9 Plugthepowersupplyi...

Страница 5: ...peatsteps1 4fortheothernostril TIP Youmayneedtotiltyourheadmoreorlesstofindthebestangleforyournose TIP YoumayneedtoadjustthelocationoftheCupforwardorbackwardtofindthe bestlocationformedicationdelivery TIP Stoppingafterthefirstfewsecondstoblowyournosemayimprovedelivery NOTE DonotturnontheCompressoruntiltheCupAssemblyisinpositioninthe nose DonotremovetheCupAssemblyfrompositionwithoutfirstturningofft...

Страница 6: ...omponents causefoggingorotherundesirableeffects Donotwashin thedishwasher Caution DonotallowwatertoentertheSupplyTubing Caution IfwaterentersorcondensationformsintheSupplyTubing connect oneendoftheSupplyTubingtotheCompressorandruntheCompressoruntilthe condensationevaporates REPLACE THE FILTER TheCompressorhasafilterinstalledintheFilterHolder Thereareadditionalfilters includedinthesystem Checkthefi...

Страница 7: ...luggedin apoweroutlet Powercordnotconnected Removeandinsert toCompressor powercordintoCompressor Poweroutletnotactive Checkthatoutlethaspower Compressor Pressingsideofpowerswitch Pressstraightdownonthe doesn tturnoff doesnotengageswitch centerofthepowerswitch ParticleSizeDistribution Performanceinformationmaynotapplytodrugssuppliedinsuspensionorhighviscosity form Insuchcases informationshouldbesou...

Страница 8: ...eRange 21 to70 C 13 to158 F RelativesHumidityRange 10to95 RH non condensing Dimension LxWxH 136x126x56mm 5 3x5 0x2 2in Weight 445g 15 6oz NASONEB Cup Assembly Dimension LxWxH 53x56x42mm 2 1x2 2x1 64in Weight 16g 0 55oz Capacity 15ml 0 5floz Phthalate Free Notmanufacturedwithanynaturallatexproducts Supply Tubing Length 1 98m 6 5ft Phthalate Free Notmanufacturedwithanynaturallatexproducts Accessorie...

Страница 9: ...essorusesnoRFenergy foritsfunction Therefore itsRFemissions areverylowandarenotlikelytocauseany interferenceinnearbyelectronicequipment 5 RF emissions CISPR11 Class B TheNasoNebCompressorissuitable foruseinallestablishments including domesticestablishmentsandthosedirectly connectedtothepubliclow voltagepower supplynetworkthatsuppliesbuildingsused fordomesticpurposes 6 Harmonic emissions EN 61000 3...

Страница 10: ...erential mode 2kVcommonmode N A Mainspowerqualityshouldbethat ofatypicalcommercialorhospital environment Voltage dips short interruptions andvoltage variationson powersupply inputlines EN 61000 4 11 0 UT 0 5cycleg At0 45 90 135 180 225 270 and315 0 UT 1cycleand 70 UT 25 30cycles at0 0 UT 250 300cycle N A Mainspowerqualityshouldbe thatofatypicalcommercialor hospitalenvironment Iftheuserof theNasoNe...

Страница 11: ...metres m b FieldstrengthsfromfixedRFtransmitters as determinedbyanelectromagneticsitesurvey a shouldbelessthanthecompliancelevelineach frequencyrange Interferencemayoccurinthe vicinityofequipmentmarked withthefollowingsymbol NOTE1 At80MHzand800MHz thehigherfrequencyrangeapplies NOTE2 Theseguidelinesmaynotapplyinallsituations Electromagneticpropagationisaffectedbyabsorptionandreflectionfromstructur...

Страница 12: ...ccordingtothemaximumoutputpowerofthe communicationsequipment Ratedmaximum outputoftransmitter W Separationdistanceaccordingtothefrequencyoftransmitter m 150kHzto80Mhz 80MHzto800MHz 800MHzto2 5GHz 0 01 l 0 12 0 23 0 1 l 0 38 0 73 1 l 1 2 2 3 10 l 3 8 7 3 100 l 12 23 Fortransmittersratedatamaximumoutputnotlistedabove therecommendedseparationdistanced inmetres m canbeestimatedusingtheequationapplicab...

Страница 13: ...RISINGFROMACOURSEOFDEALING ORUSAGEORTRADE NASONEB INC ISNOTRESPONSIBLEFORHANDLING STORAGE CLEANING ANDSTERILIZATIONOFTHEPRODUCT ASWELLASOTHERFACTORSRELATINGTOTHEPATIENT DIAGNOSIS TREATMENT MEDICALPROCEDURES ANDOTHERMATTERSBEYONDNASONEB INC S CONTROL DIRECTLYAFFECTTHEPRODUCTORTHERESULTSOBTAINEDFROMITSUSE THISLIMITED WARRANTYDOESNOTEXTENDTOANYABUSEORMISUSEOFTHEPRODUCT INCLUDING WITHOUT LIMITATION OF...

Страница 14: ...14 ...

Страница 15: ...15 ...

Страница 16: ...red in the USA Date of revision 2021 11 03 Reference Number 300 532105 01 NasoNeb Inc 153 Industrial Blvd Plattsburgh NY 12901 1 866 960 9833 1 866 565 1893 www nasoneb com MDSS GmbH Schiffgraben 41 30175 Hannover Germany Tel 49 511 6262 8630 Fax 49 511 6262 8633 EC REP ...

Отзывы: