Set Up
4
Figure 9
Figure 8
Figure 6
Figure 7
the head back around so that it is fac-
ing up and in the correct position.
Connecting the Airways
Once the head is in place, connect
the CPARLENE
®
lung bag to the right
bronchus by sliding the cap plug over
the white adapter in the bronchus.
(See Figure 6.)
Then use the provided red cap plugs
to plug the esophagus and the left
bronchus. (See Figure 7.)
This will ensure the system is sealed
and operational.
Laryngospasm
Pull back the plunger of a 20 cc
syringe and attach it to the red
stopcock. (See figure 8.) Depress the
plunger to force 1-5 cc of air into the
system. Close the tubing port at the
stopcock to maintain the pressure
within the larynx. The syringe may be
removed at this point.
Tongue Swelling
Pull back the plunger of a 20 cc
syringe and attach it to the yellow
stopcock. (See Figure 9.) Depress the
plunger to force approximately 10
cc of air into the system. Close the
tubing port at the stopcock to main-
tain the pressure. The syringe may be
removed at this point.
Caution:
Do not overinflate! Excessive
pressure may rupture the system. Stop
inflation when resistance is felt or the
larynx is observed to close. Always
release the pressure when finished
training by opening the tubing port at
the stopcock.
Now match the single keyway on the
neck front up to the slot on the back-
side of the torso, and push the head-
down until it snaps into place. Rotate