Set Up

3

    enter the empty bag.

7. Close the lower bag clamp and

shut the red plug filler cap

to the lower bag. You now

have a full vein system for

access. Elevate the blood bag.

Leaving the clamp to the

lower bag closed will allow

you to attain a flash in the

catheter as you insert the

needle into the vein. The

    higher the supply bag of blood  

    is hung, the more pressure is 

    created in the system to flash.

8. Once the catheter is installed 

    and secured, open the lower 

    bag tubing clamp to allow 

 titration of the drip in the IV 

 set-up.

9. Phlebotomy can be 

    accomplished without 

    opening the lower tubing 

    clamp.

Replacing IV Skins and Veins

Tubing and Veins: To replace 

tubing in Simulaids IV Arm or

Hand:

Steps: 

1. Remove skin and loosen 

 tubing.

PO Box 1289 – 16 Simulaids Drive   Saugerties NY 12477 USA   Tel: 845-679-2475 or 800-431-4310   Fax: 845-679-8996

www.simulaids.com   [email protected]

Page 3 of 4

IV TRAINING ARM AND HAND

Assembly & Operation Instructions

8. Once the catheter is installed and secured, open the

lower bag tubing clamp to allow titration of the drip in

the IV set-up.

9. Phlebotomy can be accomplished without opening the 

lower tubing clamp.

Tubing and Veins: To replace tubing in Simulaids IV Arm or 

Hand:
Steps: 
1. Remove skin and loosen tubing.

Replacing IV Skins and  Veins

2. Clip tubing, leaving about an inch (1") protruding 

from each hole.

3. Insert one of the supplied connectors into the end of 

each piece of tubing that is left protruding from the IV 

unit.

4. Cut the new tubing to the correct length between 

connectors.

5. Slip new tubing onto connectors.
6. Replace skin.
7. Palpate the veins to make sure they are positioned in 

the channel on the arm or hand.

Skins: To replace skins on SIMULAIDS IV Arm or Hand:
Steps: 
1. Remove the used skin from the Arm or Hand. Carefully 

cutting it off is quickest.

2. Sprinkle talcum powder from enclosed packet in the 

interior of new skin.

3. Holding so that water cannot enter the inside, dip the 

skin into hot water (120°-180°F) for approximately 15-

30 sec. This will make it easier to slip on and fit to 

inner Arm or Hand.

4. Shake out excess powder and slide the new skin over 

the Arm or Hand, pulling it into place.

5. Palpate veins to make sure they are positioned in the 

channel on the arm or hand. Adjust as needed.

6. Trim excess skin with scissors.

Item No. Description
140

IV Training Arm and Hand (1 Arm Unit/1 Hand Unit)

146

IV Training Arm

147R

New IV Training Hand Right

141R

New IV Hand Skin (Right)

141L

Original Hand Skin (Left)

142

IV Arm Skin

143

IV Arm and/or Hand Vein Set

144

IV Reservoir (1)

225

Powder to make one gallon blood

148   

WV IV Pro (4 pack)

Item numbers for IV Arm and Hand components 

are listed below:

(continued on next page)

7b.

7.

1.

4.

PO Box 1289 – 16 Simulaids Drive   Saugerties NY 12477 USA   Tel: 845-679-2475 or 800-431-4310   Fax: 845-679-8996

www.simulaids.com   [email protected]

Page 3 of 4

IV TRAINING ARM AND HAND

Assembly & Operation Instructions

8. Once the catheter is installed and secured, open the

lower bag tubing clamp to allow titration of the drip in

the IV set-up.

9. Phlebotomy can be accomplished without opening the 

lower tubing clamp.

Tubing and Veins: To replace tubing in Simulaids IV Arm or 

Hand:
Steps: 
1. Remove skin and loosen tubing.

Replacing IV Skins and  Veins

2. Clip tubing, leaving about an inch (1") protruding 

from each hole.

3. Insert one of the supplied connectors into the end of 

each piece of tubing that is left protruding from the IV 

unit.

4. Cut the new tubing to the correct length between 

connectors.

5. Slip new tubing onto connectors.
6. Replace skin.
7. Palpate the veins to make sure they are positioned in 

the channel on the arm or hand.

Skins: To replace skins on SIMULAIDS IV Arm or Hand:
Steps: 
1. Remove the used skin from the Arm or Hand. Carefully 

cutting it off is quickest.

2. Sprinkle talcum powder from enclosed packet in the 

interior of new skin.

3. Holding so that water cannot enter the inside, dip the 

skin into hot water (120°-180°F) for approximately 15-

30 sec. This will make it easier to slip on and fit to 

inner Arm or Hand.

4. Shake out excess powder and slide the new skin over 

the Arm or Hand, pulling it into place.

5. Palpate veins to make sure they are positioned in the 

channel on the arm or hand. Adjust as needed.

6. Trim excess skin with scissors.

Item No. Description
140

IV Training Arm and Hand (1 Arm Unit/1 Hand Unit)

146

IV Training Arm

147R

New IV Training Hand Right

141R

New IV Hand Skin (Right)

141L

Original Hand Skin (Left)

142

IV Arm Skin

143

IV Arm and/or Hand Vein Set

144

IV Reservoir (1)

225

Powder to make one gallon blood

148   

WV IV Pro (4 pack)

Item numbers for IV Arm and Hand components 

are listed below:

(continued on next page)

7b.

7.

1.

4.

2. Clip tubing, leaving about an 

    inch (1”) protruding from each 

    hole.

3. Insert one of the supplied 

    connectors into the end of

    each piece of tubing that is left 

    protruding from the IV unit.

4. Cut the new tubing to the 

    correct length between

    connectors.

5. Slip new tubing onto 

    connectors.

6. Replace skin.

7. Palpate the veins to make sure

    they are positioned in the 

    channel on the arm or hand.

Skins: To replace skins on Nasco 

Healthcare IV Arm or Hand:

Steps: 

1. Remove the used skin from the

    Arm or Hand. Carefully

    cutting it off is quickest.

2. Sprinkle talcum powder from 

    enclosed packet in the interior 

    of new skin.

3. Holding so that water cannot

    enter the inside, dip the

    skin into hot water (120  -180  F) 

    for approximately 15-30 sec 

PO Box 1289 – 16 Simulaids Drive   Saugerties NY 12477 USA   Tel: 845-679-2475 or 800-431-4310   Fax: 845-679-8996

www.simulaids.com   [email protected]

Page 3 of 4

IV TRAINING ARM AND HAND

Assembly & Operation Instructions

8. Once the catheter is installed and secured, open the

lower bag tubing clamp to allow titration of the drip in

the IV set-up.

9. Phlebotomy can be accomplished without opening the 

lower tubing clamp.

Tubing and Veins: To replace tubing in Simulaids IV Arm or 

Hand:
Steps: 
1. Remove skin and loosen tubing.

Replacing IV Skins and  Veins

2. Clip tubing, leaving about an inch (1") protruding 

from each hole.

3. Insert one of the supplied connectors into the end of 

each piece of tubing that is left protruding from the IV 

unit.

4. Cut the new tubing to the correct length between 

connectors.

5. Slip new tubing onto connectors.
6. Replace skin.
7. Palpate the veins to make sure they are positioned in 

the channel on the arm or hand.

Skins: To replace skins on SIMULAIDS IV Arm or Hand:
Steps: 
1. Remove the used skin from the Arm or Hand. Carefully 

cutting it off is quickest.

2. Sprinkle talcum powder from enclosed packet in the 

interior of new skin.

3. Holding so that water cannot enter the inside, dip the 

skin into hot water (120°-180°F) for approximately 15-

30 sec. This will make it easier to slip on and fit to 

inner Arm or Hand.

4. Shake out excess powder and slide the new skin over 

the Arm or Hand, pulling it into place.

5. Palpate veins to make sure they are positioned in the 

channel on the arm or hand. Adjust as needed.

6. Trim excess skin with scissors.

Item No. Description
140

IV Training Arm and Hand (1 Arm Unit/1 Hand Unit)

146

IV Training Arm

147R

New IV Training Hand Right

141R

New IV Hand Skin (Right)

141L

Original Hand Skin (Left)

142

IV Arm Skin

143

IV Arm and/or Hand Vein Set

144

IV Reservoir (1)

225

Powder to make one gallon blood

148   

WV IV Pro (4 pack)

Item numbers for IV Arm and Hand components 

are listed below:

(continued on next page)

7b.

7.

1.

4.

PO Box 1289 – 16 Simulaids Drive   Saugerties NY 12477 USA   Tel: 845-679-2475 or 800-431-4310   Fax: 845-679-8996

www.simulaids.com   [email protected]

Page 3 of 4

IV TRAINING ARM AND HAND

Assembly & Operation Instructions

8. Once the catheter is installed and secured, open the

lower bag tubing clamp to allow titration of the drip in

the IV set-up.

9. Phlebotomy can be accomplished without opening the 

lower tubing clamp.

Tubing and Veins: To replace tubing in Simulaids IV Arm or 

Hand:
Steps: 
1. Remove skin and loosen tubing.

Replacing IV Skins and  Veins

2. Clip tubing, leaving about an inch (1") protruding 

from each hole.

3. Insert one of the supplied connectors into the end of 

each piece of tubing that is left protruding from the IV 

unit.

4. Cut the new tubing to the correct length between 

connectors.

5. Slip new tubing onto connectors.
6. Replace skin.
7. Palpate the veins to make sure they are positioned in 

the channel on the arm or hand.

Skins: To replace skins on SIMULAIDS IV Arm or Hand:
Steps: 
1. Remove the used skin from the Arm or Hand. Carefully 

cutting it off is quickest.

2. Sprinkle talcum powder from enclosed packet in the 

interior of new skin.

3. Holding so that water cannot enter the inside, dip the 

skin into hot water (120°-180°F) for approximately 15-

30 sec. This will make it easier to slip on and fit to 

inner Arm or Hand.

4. Shake out excess powder and slide the new skin over 

the Arm or Hand, pulling it into place.

5. Palpate veins to make sure they are positioned in the 

channel on the arm or hand. Adjust as needed.

6. Trim excess skin with scissors.

Item No. Description
140

IV Training Arm and Hand (1 Arm Unit/1 Hand Unit)

146

IV Training Arm

147R

New IV Training Hand Right

141R

New IV Hand Skin (Right)

141L

Original Hand Skin (Left)

142

IV Arm Skin

143

IV Arm and/or Hand Vein Set

144

IV Reservoir (1)

225

Powder to make one gallon blood

148   

WV IV Pro (4 pack)

Item numbers for IV Arm and Hand components 

are listed below:

(continued on next page)

7b.

7.

1.

4.