46/106
7.c.5.
Grid ON / OFF
:
When checked, one-
minute disclosure screen will be display with grid
7.c.6.
Mark R-R interval
:
If the HR is not
correctly calculated, then the R-R intervals can be marked
so that the HR is recalculated. For marking the R-R interval,
press the ‘
Mark R-R interval
’ button, a vertical line cursor
will be displayed. Place the cursor on the R wave and click.
A second vertical line cursor will be displayed. As this cursor
is moved on the screen the value of R-R interval in
milliseconds will be displayed on the screen. Place the
cursor on successive R wave and click again. The HR for
that page will be recalculated according to the marked R-R
interval.
7.c.7.
Standard Events:
This control has the same
functionality as described before.
7.c.8.
HR edit box,
If the HR is not correctly calculated, count the QRS complex
on one minute page and enter the value in above box and
press ‘tab’ key on keyboard, changed HR value is displayed
on one min page.
7.c.9.
Beat count
:
“Beat count” button is added
on One Minute Disclosure page to calculate HR
automatically based on actual bit count in that minute.
By pressing “beat count” button on one minute page
software calculate HR, based on actual bit in that
minute.
Содержание Reveala3 XT
Страница 8: ......
Страница 57: ...57 106 Fig 7 2 Summary report Part 1 ...
Страница 68: ...68 106 Fig 7 16 Poincare Graph and Normal RR histogram ...
Страница 73: ...73 99 Reveala 3 with keyboard unit photo USB Data Cable Patient Cable ...
Страница 80: ...80 106 Click Next Click Yes Setup ...
Страница 81: ...81 106 Click Next Click Next ...
Страница 84: ...84 106 ...
Страница 86: ...86 106 By examining the device manager located in control panel then select the Hardware tab and click device ...
Страница 99: ...99 106 Name Signature Stamp Date ...