Setting the resolution / EMC bandwidth (RBW / EBW)
You selected the desired view.
To set the resolution / EMC bandwidth:
1.
Tap the
RBW
/
EBW
button.
2.
Enter the required value and select a unit.
The setting is accepted immediately and the dropdown list is closed.
Settings in the context menu
You selected the desired view.
⇒
Long tap the
RBW
/
EBW
button.
The context menu opens.
Enabling/disabling the auto resolution bandwidth (Auto RBW /
Auto EBW)
Enable or disable automatic calculation of resolution bandwidth.
To set Auto RBW:
1.
Tap the
Auto RBW
button to switch function
On
/
Off
.
2.
Press the
Esc
key to close the context menu.
Setting the resolution bandwidth entry mode (RBW Entry / RBW Entry)
This button allows you to select the entry type
Absolute
or
Relative
.
To set RBW Entry:
1.
Tap the
RBW Entry
button to select the required type.
2.
Press the
Esc
key to close the context menu.
Setting the measurement time (Meas. Time)
Here you can set the measurement time. The measurement or detector time defines the time
range used by the internal detectors to calculate their values (
+Pk
,
RMS
,
-Pk
, ...).
To set the measurement time:
You selected the desired view.
1.
Tap the
Meas. Time
button.
2.
Enter the required value and select a unit.
The setting is accepted immediately and the dropdown list is closed.
Configuring the measurement performance
(Config)
To change the configuration:
1.
Tap the header of the desired view.
2.
Tap the
Config
button.