background image

12

11.1  To test, let vessel cool for 10 minutes if hot.
Refill vessel to proper water level, turn Function
Switch to Dryer 'OFF' position, turn Timer and
Power Switch 'ON' and rotate Autoclave
Temperature Control to highest setting. If the
Temperature Pilot Light is 'OFF' connect an AC
Voltmeter across T.B. Terminal #2 and T.B.
Terminal #4. Full line voltage should be present. If
not, see Sections #3, #4, #5 and #6. Now move
meter test lead from T.B. Terminal #4 to T.B.
Terminal #5. Full line voltage should be present
while vessel is in  its' initial heating stage before
pressurization. If line voltage is not present at T.B.
Terminal #5, replace Autoclave Temperature
Control. See Section 17. If this is a new control,
ensure one wire is connected to 'Common' and
the other wire connected to the 'N.C.' (normally
closed) contact on the controller.

11.2  Allow vessel to continue heating until
maximum temperature and pressure are
readched, If pressure goes above 30 PSI or
controls at less tan 27 PSI at maximum setting,
go to Section 12 to attempt adjustment. If
controlled pressure fluctuates more than 2 PSI
from set point after one hour, replace Autoclave
Temperature Control. See Section 17.

12.0  AUTOCLAVE TEMPERATURE CONTROL
ADJUSTMENT:

12.1  While vessel is under pressure in autoclave
cycle, loosen the two set screws in the control knob
and remove knob. Set the control knob shaft so
that the Autoclave Temperatuer Conteol regulates
at 27 PSI ± 1 PSI/133°C±1°C. If pressure/
temperature is high, turn shaft clockwise. Allow
controller ample time to reach set point after
making an adjustment. If pressure varies over 2
PSI or fluctuates inconsistently, the control should
be replaced. See Section 17.

12.2  Once the control is properly adjusted at 27
PSI ± 1 PSI, place control knob on shaft. Turn
knob counterclockwise untiil knob reaches stop.
Now tighten knob to control shaft.

12.3  Turn Control Knob fully clockwise. allos
ample timer for vessel to cool and be controlled at
low set point. Pressure should drop to at least 15
PSI and control ±1 PSI. If Controller will not do
this, it should be replaced.

13.0  HEATING ELEMENTS:  Autoclave and Dryer
Elements. Unplug unit from electrical outlet.

13.1  To test Dryer Element (Top Element), remove
both wires from the Dryer Element terminals on
top of the vessel. Connect an Ohmmeter across
the two element terminals. The Dryer Element
should measure 11.5 ohms ± 0.6 ohms on 120V
Elements or 46 ohms ± 3 ohms on 240V Elements,
if not, replace Element. See Section 18. Now
conduct test in Section 13.3.

13.2  To test Autoclave Heating Element (bottom
Element), remove bottom access cover and
remove both wires fromthe Heating Element
terminals. Connect an Ohmmeter across the two
Element terminals. The autoclave element should
measure 10.3 ohms ± .6 ohms on 120V Elements
or 41 ohms ± 3 ohms on 240V Elements. If not,
replace Element. See Section 18. Now conduct
test in Section 13.3.

13.3  Now select highest resistance range on
Ohmmeter. Touch one test lead to either terminal
on Element on touch other test lead to the vessel
body. This test will give an indication whether or
not the Element is shorting to ground. Do not touch
ends of test leads with hands. If you get any meter
indication less than infinity, replace Heating
Element. See Section 18. Replace wires to
Element when test is completed.

14.0  DRYER TEMPERATURE CONTROL: This
is an infinite control which varies the 'ON-OFF' rate
and current 'ON' time which is being supplied to
the Dryer Heating Element. The rate/time interval
is a function of the Dryer Control Setting.

14.1  The Dryer Temperature Control has 5 wire
connections labeled; P, L-1, L-2, H-1 and H-2.
Connections L-1 and L-2 are the power input point.
Connections  H-1 and H-2 are the power output
points which lead to the Dryer Element. Connection
P supplies power to the Dryer Pilot Light when the
Dryer Temperature Control is turned ON.

14.2  To test Dryer Temperature Control first test
Dryer Heating Element as in Section 13. Now turn
Timer, Power Switch and Function Switch 'ON'.
Connect an AC Voltmeter across the two Dryer
Temperature Control contacts labeled 'L-1' and
'L-2' to ensure power is present at Controller
inputs. If line voltage is not present, see Sections
3, 4, 5 and 6.

Содержание 8100-TD

Страница 1: ...ster VA USA Phone 540 869 9892 Toll Free 800 621 8820 FAX 540 869 0130 Manual P N 34000716 Rev D Dated 23AUG99 MODEL 8100 TD TEST CHAMBER E SERIES POWER ON VALVE 30 CONDENSATE DRAIN 40 20 OFF HOURS TI...

Страница 2: ...KES NO REPRESENTATIONS OR WARRANTIES WITH RESPECT TOTHISMANUAL INNOEVENTSHALLNAPCOBELIABLEFORANYDAMAGES DIRECT OR INCIDENTAL ARISING OUT OF OR RELATED TO THE USE OF THIS MANUAL For repair information...

Страница 3: ...INDEX DATE REVISION STATUS Initial release A AMENDED PAGES NOTES D 8 99 All Update part no s and combined service manual 3400745 with user manual...

Страница 4: ...ding The Test Chamber 3 Procedure for Operation of Test Chamber 3 Procedure for Operation of Dryer Cycle 4 Maintenance Suggestions 4 Sterilizing Cycle Data 5 Explanation of System Operation 6 Troubles...

Страница 5: ...cted before leaving our factory Save all packing material if apparatus is received damaged Responsibility for safe delivery was assumed by the carrier upon acceptance of the shipment therefore claims...

Страница 6: ...6 Condensate valve and drain Used to drain excess water from condensation tank after every test chamber cycle INITIAL PROCEDURE AFTER UNPACKING UNIT WARNING FOR PERSONAL SAFETY THIS APPARATUS MUST BE...

Страница 7: ...ol correspond to higher temperature settings The point reference low on the dial face correspond to a temperature of 121 C 250 F Numbers between the reference correspond to temperatures within the ran...

Страница 8: ...witch to ON position set the dryer temperature control to LOW position set the timer to desired drying time interval set the power switch to ON position At the end of the set time interval an alarm wi...

Страница 9: ...8 10 LIQUIDS MEDIUM 121 123 250 254 15 15 NOTE Sterilizing time will vary with the quanitity of items in the Autoclave vessel Use a steam clox sterilization indicator with each load The data shown ab...

Страница 10: ...and a pressure relief valve The low water cut off switch will be activated if the unit is allowed to run out of wate caused by not filling vessel to water indicator at the beginning of each cycle or...

Страница 11: ...open See Section 11 3 Unit does not heat Tempera ture Light OFF Timer Light ON 1 Autoclave Temperature control See Section 11 2 Pressure leak See section 15 4 Operating Temperature and pressure cannot...

Страница 12: ...ons 6 See Problems 7 9 7 Safety pressure relief valve seat leaking See Section 15 1 Dryer Element open See Section 13 11 Drying element does not heat Dryer Light ON Timer Light ON 12 Drying element do...

Страница 13: ...se connections in the power plug 3 0 POWER SWITCH Unplug unit from electrical outlet 3 1 With Power Switch in OFF position measure the resistance across the two white wires Then measure across the two...

Страница 14: ...o reach this temperature from ambient Observe temperature gauge on door while unit is heating Meter should display a change from zero volts to full line voltage when temperature gauge reaches 103 C 1...

Страница 15: ...il If you cannot hear the Solenoid Valve actuate or if it clatters when energized it should be replaced Unplug the unit and return the Solenoid wire from T B Terminal 3 to T B Terminal R6 10 2 Now tes...

Страница 16: ...15 PSI and control 1 PSI If Controller will not do this it should be replaced 13 0 HEATING ELEMENTS Autoclave and Dryer Elements Unplug unit from electrical outlet 13 1 To test Dryer Element Top Eleme...

Страница 17: ...Use a second tool to hold part from turning 15 5 After repairing a leak always retest unit in Autoclave Cycle If a repaired fitting leaks after re testing it should be replaced Order the preformed cop...

Страница 18: ...d start Autoclave Cycle Now proceed with Autoclave Temperature Control Adjustment per Section 12 Also check for fitting leaks while unit is under pressure 18 0 AUTOCLAVE AND DRYER ELEMENT REPLACEMENT...

Страница 19: ...ed Lime Citric Acid boiling Dakin s Solution Ferrous Chloride Mercury Salts LysolMercuric Chloride Phenol Potassium Permanganate Stanous Chloride Sodium Hypochlorite Tartaric Acid Potassium Thiocyanat...

Страница 20: ...16 POWER SWITCH FUNCTION SWITCH DRYER AUTOCLAVE DOOR GASKET WATER BAFFLE SHELF DRAIN VALVE AND DRAIN 60 HOUR TIMER INDICATOR LAMPS AUTOCLAVE TEMP CONTROL DRYER TEMP CONTROL...

Страница 21: ...RYER TEMP CONTROL AUTOCLVAE TEMP CONTROL DRYER ELEMENT TERMINALS SOLENOID VALVE CONDENSOR TANK PRESSURE RELIEF VALVE BUZZER FUSE HOLDER POWER CORD SENSOR ASSEMBLY RELAY TERMINAL BLOCK TIMER DRAIN VALV...

Страница 22: ...e T 8 4 0 5 4 2 1 5 t i K t n e m e l E g n i t a e H 7 2 1 5 4 2 1 5 5 3 5 1 0 0 4 3 t i K t n e m e l E g n i y r D 5 4 0 5 4 2 1 5 6 4 0 5 4 2 1 5 f f o t u C l a m r e h T r e y r D 4 9 1 9 3 2 4...

Страница 23: ...T r e s n e d n o C 5 8 1 1 0 0 4 3 e v l a V f e i l e R e r u s s e r P 6 1 1 9 6 2 4 3 y l b m e s s A e v l a V n i a r D 5 9 6 1 4 5 4 3 l e s s e V o t r o s n e S p m e T e b u T 5 2 9 2 0 0 4...

Страница 24: ...L 1 L 2 P GROUND 40 BLACK BLACK AUTOCLAVE LOW WATER CUT OFF SWITCH OPEN 132 C CLOSE 110 C BLACK BLACK WHITE BLACK COM BLACK COM 13 9 31 BLACK BLACK 9 10 DRYER FUNCTION SW 30 DRYER CONTROL WHITE OPEN...

Страница 25: ...T BE LIABLE FOR CONSEQUENTIAL DAMAGE RESULTING FROM THE USE OF ITS PRODUCTS PRECISION will not assume responsibility for unauthorized repairs or failure as a result of unauthorized productmodification...

Отзывы:

Похожие инструкции для 8100-TD