RESISTANCE AND CONTINUITY MEASUREMENTS 

1. 

Insert 

Black test lead

 into the 

negative COM jack

 and 

Red test lead

 into the 

Positive 

(V) jack. 

2. 

Set the function switch to the         

•))) 

 position.                                                              

3. 

Use the multifunction 

MODE

 button to select resistance. 

4. 

Touch test probe tips across the circuit or component under  
test. 

IMPORTANT

:  It is best to disconnect one side of the device under test so  

the rest of the circuit will not interfere with the resistance reading. 

5. 

For Resistance tests, read the resistance on the LCD display. 

6. 

For Continuity tests, if the resistance is < 100

, a tone will sound. 

DIODE MEASUREMENTS 

1.  Insert 

Black test lead

 banana plug into the 

negative COM jack

 and 

Red test lead

 

banana plug into the 

positive diode jack

2.  Turn the rotary switch to the       

•)))

 position. 

3.  Press the 

MODE

 

button until “     “ appears in the display. 

4.  Touch the test probes to the diode under test. Forward voltage will indicate 0.4V to 0.7V. 

Reverse voltage will indicate “

OL

”. Shorted devices will indicate near 0mV and an open 

device will indicate “

OL

” in both polarities. 

 

 

 

CAPACITANCE MEASUREMENTS 

WARNING:

 To avoid electric shock, disconnect power to the unit under test and 

discharge all capacitors before taking any capacitance measurements. Remove the 
batteries and unplug the line cords. 

1. Set the Rotary Function switch to the cap position. 
2. Insert 

Black test lead

 banana plug into the 

negative COM jack

    Insert 

Red test lead

 banana plug into the 

positive (V) jack

3. Touch the test leads to the capacitor to be tested. 
4. Read the capacitance value in the display 

FREQUENCY AND DUTY CYCLE % MEASUREMENTS 

1.  Set the function switch to the 

(V)

 position. 

2.  Insert the 

Black lead banana plug 

into the 

negative COM jack

 and the 

Red lead

 

banana plug

 into the 

positive (V) jack

3.  Select Hz or % duty with the (

Hz

/%

)

 button. 

4.  Touch the test probe tips to the circuit under test. 

5.

 

Read the frequency on the display.

 

 

TEMPERATURE MEASUREMENTS 

WARNING:

To avoid electric shock, disconnect both test probes from any source of        

  voltage before making a temperature measurement. 

1. 

  Set the function switch to 

TEMP 

2. 

  Insert Temperature Probe into 

negative (COM) and (V) jacks

, making    

 

  sure to observe the correct polarity 

3. 

  Select 

C or 

F with the 

MODE

 button 

4. 

  Touch Temperature Probe head to the surface area you wish to measure.  

 

  Maintain constant contact with the part under test until the reading stabilizes  

 

  (approx. 30 seconds) 

5. 

  Read the temperature in the display. The digital reading will indicate the proper  

 

  decimal point and value 

 

 

WARNING: 

To avoid electric shock, be sure the thermocouple has been 

 

 

removed before changing to another measurement function 

DATA HOLD AND BACKLIGHT 

To freeze the LCD meter reading, press the data hold button. The Data Hold button is 
located on the left side of the meter (top button). While data hold is active, the 

HOLD

 

display icon appears on the LCD. Press the data hold button again to return to normal 
operation. 

Note: The HOLD feature will activate when the Backlight is turned on. Press the HOLD key again 
to exit Hold. 

The backlight function illuminates the display and is used when the ambient light to too low 
to permit viewing of the displayed readings. Press the     (HOLD) button for one second to 
turn the backlight on and press the button a second time to turn the backlight off. 

MANUAL RANGING FEATURE 

Auto-

Ranging is meter’s default setting.  To deactivate press the 

Range

 button to go to 

manual ranging. Each press of the range button will step to the next range as indicated by 
the units and decimal point location.  Press and hold the 

Range

 button for two seconds to 

return to Auto-Ranging.  NOTE:  Manual ranging does not function in the AC Current, 
Diode, and Continuity check functions 

BATTERY REPLACEMENT 

1.  Remove Phillips head screw from back of meter 
2.  Open battery compartment 
3.  Replace dead battery with one 

“9V” battery (NEDA1604, 6F22 006P)  

4.  Re-assemble the meter 
 

TECHNICAL HOTLINE: 

888-227-1603 

For any technical assistance or possible service issues, please call the toll free 
number above.  Especially if you feel the need to return your instrument.  Please 
call us first and let us help you!

 

 

WARRANTY INFORMATION 

This product is warranted to be free from manufacturer defects for ONE YEAR from date of 

purchase, and carries a One Year “Repair/Return” warranty (code 724).  If this product fails 

during the first 12 months due to faulty materials or workmanship, it will be repaired or 

replaced free of charge, at the discretion of the manufacturer. 

NOTE:  This one year warranty does not cover dead batteries and blown fuses.

 

IMPORTANT REPAIR/RETURN INSTRUCTIONS

Users must call and request Return Authorization # prior to sending any product in for out-

of-warranty repair by calling 888-227-1603.

  

Out of warranty repairs billable to the end user. 

For warranty and service coverage after YEAR ONE please call 888-227-1603. 

REPAIRS:  MAIL TESTER FREIGHT PREPAID TO: 

  NAPA Meter Repairs  /  139 Elizabeth Ln.  /  Genoa City, WI 53128 

NOTE: 

RETURNS AUTHORIZATION and BRIEF DESCRIPTION OF PROBLEM MUST ACCOMPANY 

ALL REPAIR RETURNS 

Red               Black     Black           Red 
Probe            Probe    Probe           Probe 

Forward test        Reverse test

 