page
36
A/S manual
Check the water supply
(DES-10K)
3. Troubleshooting >
Check the saline /
Check the water supply
/ Check the other errors
※ Do not touch, modify or repair the parts inside the device.
※ In case of abnormal operation, unplug the power cord and request a repair by contacting the seller.
Action
Cause
-When the Water supply checking lamp
is blinking with a buzzer alert.
-When the saline solution is not supplied
normally because of impurities in the
Water valve.
-When the water flows out from the
Disinfectant outlet hose, even if
the power is shut off.
-Please check the Water supply valve
①
is opened properly.
-Clean or exchange the Water supply solenoid valve.
(Refer to the next page > How to clean the Water supply solenoid.)
<OPEN>
Supply
of water
Direction of Salt tank
1
2
Supply
of water
<CLOSE>
Direction of Salt tank
1
2