A
PPROACHING
THE
SAMPLE
53
- If you are operating in Static Force mode, select a CONTR type canti-
lever.
In dynamic mode, the instrument will automatically determine the vibra-
tion frequency used in the dynamic mode. To do this, it uses a measurement
of the cantilever vibration amplitude as a function of excitation frequency.
This measurement is not displayed when using the default settings. Never-
theless, it is instructive to see this measurement at least once. To display the
measurement:
- Open the User interface dialog via the menu ‘Options/Config User Inter-
face...’.
- Select the ‘Standard level’ user interface mode.
- Enable ‘Display sweep chart’ in the Mode Properties section of the Oper-
ating mode panel.
- Click the ‘Set’ button.
The software will now perform a coarse and fine frequency sweep, and dis-
play the results (figure
). For more details on the fre-
quency search procedure see section
Now restore the default settings:
- Disable ‘Display sweep chart’ in the Mode Properties section of the Oper-
ating mode panel.
- Select the ‘Easy level’ user interface mode.
Fine frequency sweep: Cantilever vibration amplitude as a function of frequency.
Содержание easyScan 2 AFM
Страница 1: ...Operating Instructions easyScan 2 AFM Version 1 6...
Страница 7: ...7...
Страница 86: ...AFM THEORY 86 Scanner coordinate system x y...
Страница 159: ...THE SCRIPT CONFIGURATION DIALOG 159 This page was intentionally left blank...
Страница 160: ...AUTOMATING MEASUREMENT TASKS 160 This page was intentionally left blank...
Страница 161: ...THE SCRIPT CONFIGURATION DIALOG 161 This page was intentionally left blank...
Страница 163: ...163 Window Operating windows Imaging 120 Positioning 115 Report 153 Spectroscopy 128...
Страница 164: ......