Trillium 240
Seismometer
Insulating Cover
Страница 1: ...Trillium 240 Seismometer Insulating Cover ...
Страница 2: ...This helps attenuate temperature variations conducted by the cable If your installation requires a different exit point for the sensor cable you can separate the centre plug from the outer main gasket You can then rotate the outer gasket to move the cable outlet as required A modified instal lation provides less insulation for the sensor cable Installing the Insulating Cover 1 Remove the pre cut g...
Страница 3: ...ely centred between the edge of the gasket and the centre plug line 8 Look inside the cover to orient the connector groove to the sensor connector Place the cover over the sensor Caution It is very important that the cover and gasket not touch the sensor or cable connector A continuous air gap is required for the Insulating Cover to work properly 9 Position the cover so that it will not be in cont...
Страница 4: ...nanometrics ca www nanometrics ca Dimensions Support If you need technical support please submit your request via the Nanometrics customer support site or by email or fax Include a full explanation of the problem 168 7 1 29 5 2 0 6 6 7 75 80 6 6020 7 5 168 7 1 29 5 0 16 216 1 1 6 PP Customer support site http support nanometrics ca Email techsupport nanometrics ca Fax To Support at 1 613 592 5929 ...