Troubleshooting
· Check on/off switch on back side of unit.
· Check power source or contact your distributor.
· Touch the LCD screen with your finger .
· Check if power supply is connected and
power switch is on.
· Reset the power button.
· Check the connection between scope and station.
· Connect connection cable to the 2
nd
scope unit after
connecting to the 1
st
scope unit completely.
· Check whether the focus interface is inactive or not.
· Move the focus knob to the opposite side.
· Re-optimize the brightness and exposure value.
· Reset the power button.
· Swipe the stage with cotton swab carefully.
· Eliminate any dust on culture dish.
· Check for and remove any condensation on the
lid of the culture dish.
·
Warm up the instrument during 1 hour before
monitoring.
(power should be on when warming up
the instrument)
·
Cell culture flask should be wetted by the
culture media before making movie.
· Eliminate dust on a culture dish.
· Check for and remove any condensation on the lid
of the culture dish.
Installation
JuLI
TM
Br does not power up
LCD screen is black
Focusing and monitoring
menu are not activated
Get an error message
during the connecting the
2
nd
scope unit
Focus
Focus knob or focus
interface doesn’t work
Monitoring
Poor image
(Too dark and too bright)
Dirty image
Time-lapse images become
dark and bright
JuLI
TM
Br, Cell History Recorder ©2012 NanoEnTek Inc.
53