background image

ADAM-rWBC2, residual Leukocyte Counter         

13

General Operation

Calibration

Calibration test 

Prior to use ADAM-rWBC2, calibration test MUST be performed. The calibration test 
should be performed to verify if optic and auxiliary structures of ADAM-rWBC2 are com-
pletely ready for the real sample test by using Standard bead solution provided in ADAM-
rWBC kit.

The device is ready to use ONLY if the calibration result is within range as indicated on the 
individual Standard bead solution provided.

Refer to a package insert of ADAM-rWBC Kit for specific instruction of use.

Leukocyte-reduced samples

Samples should be tested within 48 hours after leukocyte reduction.

A sample is mixed with r-Solution for staining. After staining, the sample is loaded into r-
Slide and left to be settled. After settling, r-Slide is inserted into ADAM-rWBC2 and count-
ed. Refer to a package insert of ADAM-rWBC Kit for specific instruction of use.

The ADAM-rWBC2 must be calibrated after each start up and at least once each day of use. 
    1.  Let beads equilibrate to room temperature before use (~10 minutes). 
    2.  Mix by rocking gently, including upside down.
    3.  Load 100 µL of the beads into the r-Slide. Let settle for 40-60 seconds.
    4.  Insert loaded r-Slide into ADAM-rWBC2 main device and press RUN SAMPLE            to start.

WARNING

The instrument does not detect incorrectly diluted samples. 
Please refer to ADAM-rWBC Kit package insert.

!

CAUTION

Avoid bubbles which may negatively affect the result.
Do not reuse r-Slide. Reusing r-Slide may result in errors.
Do not use expired r-Solution, r-Slide or Standard bead solution.
Store r-Solution at 2~8˚C. The expiration date is one year. 
Do not use after the expiration date shown on the label. Opened r-Solution 
may be stored at 2~8˚C and used for up to 6 months, or until the expiration 
date.

!

Sample Preparation 
and Testing

Содержание ADAM-rWBC2

Страница 1: ...ADAM rWBC2 residual Leukocyte Counter 1 Instruction Manual...

Страница 2: ...82 2 6220 7942 Fax 82 2 6220 7999 NanoEnTek America Inc 240 Bear Hill Road Suite 101 Waltham MA 02451 USA Tel 1 781 472 2558 Fax 1 781 790 5649 The information in this manual is described as accuratel...

Страница 3: ...ADAM rWBC2 System Installation Environmental Requirements Power On and Initial Display General Operation Calibration Sample Preparation and Testing Run Sample Result Analysis Lock Power off Maintenan...

Страница 4: ...leukocytes in a blood component for transfusion Targeting the leukocyte ADAM rWBC2 uses fluorescent dye to perform cell dyeing and automated cell counting Automated cell counting elimi nates user bia...

Страница 5: ...d very user technique dependent NanoEnTek has developed ADAM rWBC2 which is based on fluorescent microscopy techniques for counting cells ADAM rWBC2 utilizes sensitive fluorescence dye stain ing LED o...

Страница 6: ...mples white blood cells with DNA containing nuclei are stained while platelets and red blood cells without nuclei are not stained The ADAM rWBC Kit consists of r Solution and the disposable 1 test r S...

Страница 7: ...vdst nanoentek com Any damage incurred during transit must be filed with the carrier Quantity Main device 1 ADAM rWBC Kit Starter Kit with Package Insert 1 Instruction manual 1 Adapter Power cord 1 In...

Страница 8: ...ions as unload Protects the alignment of stage from external shock when ADAM rWBC2 is being moved Lock ADAM rWBC2 before turning it off or moving it Performs all procedures of automatic counting Slide...

Страница 9: ...rWBC2 Product Description Description USB Port 1 2 3 PC port Port for software and firmware updates Connects ADAM rWBC2 power cord to wall outlet Connects USB cable to ADAM rWBC2 and PC for using a s...

Страница 10: ...r airborne particles Allow for 10 cm 4 inches minimum space around the device for proper airflow Do not place any objects on the device 1 Check the connection of ADAM rWBC2 and power cord 2 Press the...

Страница 11: ...ff main power and restart device If this message still appears after restarting contact your local distributor or ivdst nanoentek com System State Error Messages during booting Cause 0x00000C00 0x0000...

Страница 12: ...s when turning off the device without locking If this message still appears after restarting contact your local distributor or ivdst nanoentek com Device unlocked CAUTION When you turn off the device...

Страница 13: ...settling r Slide is inserted into ADAM rWBC2 and count ed Refer to a package insert of ADAM rWBC Kit for specific instruction of use The ADAM rWBC2 must be calibrated after each start up and at least...

Страница 14: ...Operation Please check the procedure of sample preparation and testing below For more detailed information please refer to the next page 3 Result 2 Insert r Slide 1 Mix sample with r Solution Load the...

Страница 15: ...tube well 100 L sample 400 L reagent 5 Load 100 L mixed sample reagent onto a r Slide 6 Wait 4 7 minutes for sample settling 7 Insert r Slide into ADAM instrument and press RUN SAMPLE to start 8 After...

Страница 16: ...ion and Testing CAUTION Avoidbubbleswhichmaynegativelyaffecttheresult WARNING Sampleloadingerror Becautiousofloadingthecorrectvolume 100 L ofthesampleintor Slide Theinstrumentwillnotdetectloworhighsam...

Страница 17: ...ction of the arrow on the slide The instrument will not detect if slides are inserted incorrectly See pictures belowforproperinsertion CAUTION Pleaseinsertorremovether slidewhentheslideholderisfullyej...

Страница 18: ...s by pressing RUN SAMPLE It may take about 1 minute longer for auto focus at the initial test The total number of leukocytes will be displayed after being automatically calculated by ADAM rWBC2 softwa...

Страница 19: ...of more than 100 pixels or the number of cells is more than 160 in 1 frame this is considered as error frame If these error frames are more than 5 then the error message Bad Sample will be displayed o...

Страница 20: ...ce If there is no operation for 3 minutes the lock function will be activated automatically When the device is locked the screen will be changed as shown below If you press the power button for 2 3 se...

Страница 21: ...C2 does not need regularly scheduled maintenance because ADAM rWBC2 has no consumable materials to replace Please clean the exposed surface of ADAM rWBC2 frequently or before testing using a soft clot...

Страница 22: ...ower cord Problem with Windows OS Check power source Replace or Repurchase Unplug the adapter and plug it back in to power on the device Booting error message Result error message System state System...

Страница 23: ...a repair instrument It is recom mend that you save the original packing material in which the instrument was shipped This warranty should be limited to the replacement of defective products For any i...

Страница 24: ...r Slide Operating environment condition Temperature Humidity Altitude Transportation storage environment condition Temperature Humidity 0 C Temperature 40 C 10 Humidity 90 Altitude 2 000 m 30 C Temper...

Страница 25: ...vice 1 Barcode scanner 1 r Solution 25mL 1 Standard bead solution 7mL 1 Accessories Cat No Product Contents Quantity Each component and power cord are available for purchase separately Please contact...

Страница 26: ...al function and damage Never insert any objects especially metallic into the air vents of the instrument as this could result in electrical shock personal injury and equipment damage To avoid a potent...

Страница 27: ...de l instrument N inserez jamais aucun objet dans les orifices d aeration de l instrument car cela pourrait entrainer un choc electrique des blessures chez les utilisateurs et des dommages d equipeme...

Страница 28: ...certified by TUV Product Services to be in conformance with the applicable safety standard for the US and Canada LED USB Connection In vitro diagnostic medical device Disposal of your old appliance Wh...

Страница 29: ...se There are no adjustable components inside the instrument If a malfunction is found refer to an authorized service person Do not attempt to service the equipment This manual is only available in Eng...

Страница 30: ...tion or technical assistance please call or email NanoEnTek Inc 851 14 Seohae ro Paltan myeon Hwaseong si Gyeonggi do 18531 Korea Tel 82 2 6220 7942 Fax 82 2 6220 7999 NanoEnTek America Inc 240 Bear H...

Страница 31: ...ADAM rWBC2 residual Leukocyte Counter 31...

Страница 32: ...7942 Fax 82 2 6220 7999 NESMU AR2 001E V 0 3 NanoEnTek America Inc 240 Bear Hill Road Suite 101 Waltham MA 02451 USA Tel 1 781 472 2558 Fax 1 781 790 5649 EC Representative MT Promedt Consulting GmbH...

Отзывы: