![NANOE Zetamix ZetaSinter Скачать руководство пользователя страница 5](http://html1.mh-extra.com/html/nanoe/zetamix-zetasinter/zetamix-zetasinter_operation-manual_3210285005.webp)
5
ZetaMix System
®
c.
Operation
Temperature controller
For each material reference, the heat program must be defined by the operator according to
guidelines provided by Nanoe (www.zetamix.fr).
The maximum safety temperature is set at 1600°C. Above, the temperature controller automatically
stops the heat treatment.
The recommended maximum working temperature is 1550°C.
Heating and cooling rates may not exceed 5
o
C/min (except emergency stop).
Do not turn off the power supply of the equipment if the furnace temperature is above 500°C.
After replacing the temperature controller, its settings must be adjusted before use.
Security warning
Do not touch the inner or outer surface of the furnace during or immediately after use.
Do not place any objects on the furnace body.
d.
Maintenance
The power supply must be switched off before any maintenance work.
Danger due to electrical current.
During installation and maintenance
work, the electric supply to the furnace
must be switched off.
Work on the electrical equipment may
only
be
performed
by
qualified
personnel.
Hot surfaces, danger of burning.
You may not always realize that surfaces,
such as the furnace walls are hot. Do not
touch the surface.