Naneos Partector Скачать руководство пользователя страница 1

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Partector 

Aerosol Dosimeter 

 
 
 
 
 

 

 

 

 

 

 

Operation Manual 

English 

 

 

September 2019 

 

 

 

 

 

 

 

 

 

 

 
 
 
 
 

 

 

 
 

 A nanoparticle dosimeter for easy workplace exposure monitoring

 

Содержание Partector

Страница 1: ...Partector Aerosol Dosimeter Operation Manual English September 2019 A nanoparticle dosimeter for easy workplace exposure monitoring...

Страница 2: ...www naneos ch 2 17 Document information naneos particle solutions gmbh 2019 Related device versions V2_3 V2_4 Related firmware versions 397 Document Date 2019 09 20 Document Revision K...

Страница 3: ...sulting from incorrect use and handling Safety notes Never operate the instrument at condensing conditions Do not blow into the instrument Do not open the partector as you may damage it Do not operate...

Страница 4: ...ee air Time resolution 4s standard faster as an option Airflow active 0 5 l min Battery power 10h approximately Charging via USB Data storage SD card PC Interface USB serial by virtual COM Port Dimens...

Страница 5: ...the measured data on a graphical display and stores it on a SD card A user defined LDSA alarm level can be set If the alarm level is exceeded the instrument s internal buzzer is turned on to produce a...

Страница 6: ...onductive silicone tubing This type of tubing is frequently used because with conductive tubing there are no electrostatic losses of particles and the silicone tubing is very flexible Unfortunately th...

Страница 7: ...SD USB adapter provided Recording measurements on the SD card If a SD card is inserted in the partector at startup the device automatically starts recording on the SD card when the device is turned on...

Страница 8: ...so the charge state of the battery If no key is pressed for more than one minute the display goes dark to prolong battery life By pressing any key the display turns on again The lightning symbol indic...

Страница 9: ...he LDSA display is color coded to provide a quick assessment of air quality LDSA values below 50 m2 cm3 are displayed in green values above 250 m2 cm3 in red and values in between in yellow Figure 4 c...

Страница 10: ...date of calibration Pressing the status soft button brings you to the status screen Pressing the alarm soft button brings you to the alarm screen Status screen The status screen shows the most import...

Страница 11: ...screen The statistics screen allows you to view events detected in the past By pressing and you can change the shown time axis from minute to hour and day Extended operation information Different inl...

Страница 12: ...your favourite data analysis program Excel Origin Igor etc The data file is a simple tab delimited text file File names encode the recording time and device serial no For example if you start recordin...

Страница 13: ...show soft button will display details on the warning Some error conditions may disappear again by themselves particularly such related to high humidity upon which the warning message will also disappe...

Страница 14: ...aneos Buffer overflow software loop time delayed report error to naneos You can still measure with the device if it only appears infrequently RH error relative humidity is very high and may hinder mea...

Страница 15: ...er hatch Check condition of sealing o ring Contact naneos for a replacement if it is brittle or has visible cracks Take an artist s brush and carefully slide it along the thin HV wire Too much force c...

Страница 16: ...of 89 336 EEC with amendments 2011 65 EC Restriction of hazardous substances RoHS The technology of the partector is patent pending Related documents Partector quick start guide www naneos ch pdf par...

Страница 17: ...www naneos ch 17 17 Customer service contact naneos particle solutions gmbh Dorfstrasse 69 5210 Windisch Switzerland Mail info naneos ch Phone 41 56 560 20 70 www naneos ch...