Nalco TRASAR 2 Скачать руководство пользователя страница 2

NAMPAC

May 1, 2008

Page 2

8.

Rinse and dump the well three times with TRASAR Calibration Standard 
S0920.

9.

After dumping the third rinse, fill the well with S0920 calibration standard.

10. Press the “READ/ENT” button
11. The screen will blink “MEAS” several times (for about 10-20 seconds).
12. If the calibration is successful, the display will show a reading for the 

calibration solution of between 190 and 210 ppm, and you are now ready to 
begin testing.  If the solution reads outside this range, please repeat the 
calibration procedure.

Procedure for testing the Pen Fluorometer Calibration

1.

Make sure plug is in bottom of pen fluorometer, hold pen horizontally.

2.

Rinse and dump the well three times with S0920 calibration solution.

3.

After dumping the third rinse, fill the well with S0920 solution.

4.

Press “READ/ENT” on controller.

5.

The screen will blink “MEAS - - - -”.  After a few seconds, a reading will 
appear.

6.

If the Pen Fluorometer is calibrated properly, the display will read between 
190 and 210 ppm.  If the reading is outside this range, recalibrate the pen.

TRAC107 Test Procedure

1.

Collect the chill water system sample that you would like to measure.

2.

Fill 25ML graduated cylinder to the 5ML mark with chill water sample.  Fill 
to the 20ML line with Distilled Water (using syringe for distilled water will 
simplify this step) and mix sample in graduated cylinder

3.

Make sure plug is in bottom of pen fluorometer, hold pen horizontally.

4.

Rinse and dump sample in the well three times with sample from step #2.

5.

After dumping the third rinse, fill the well with sample from step #2.

6.

Press “READ/ENT” on controller.

7.

The screen will blink “MEAS - - - -”.  After a few seconds, a reading will 
appear.

8.

Note the reading on the screen from step #7.

9.

Multiply the reading from step#7 by 40.  This is your TRAC107 reading in 
ppm.  The recommended range for TRAC107 is 2000-2200 ppm.