Keep the uniformity and homogeneity of the liquid to be measured
d
.
To perform measurements, the rotor should be placed at the centre of
the recipient and the protective frame should properly placed
e
.
Make sure the rotor is perfectly clean
f
.
Turn the equipment off when changing from high to low speed or keep it
for a longer time at low speed to avoid the error caused by the liquid inertia.
g
.
When working with low viscosity, use rotor nº 1 and when working at high
viscosity use rotor nº 4.
h
.
The measure time for the viscosity measured at low speed is relatively
longer.
i
.
Raise the viscosimeter by turning the knob of the ascending and falling
rack every time the rotor or sample is going to be changed.