14
Y-KH15-007A Rev.3
8. WIRING AND CONNECTION
■
8.1 Wiring
1) Refer to the table below to obtain a suitable shielded cable.
Model number
Terminal opening size
Gasket inner dia.
(mm)
Suitable cable dia.
(mm)
KH15-J-
□□
terminal box
(small)
G3/8
10a
(standard)
φ7
φ6 to 7
10b
φ8
φ7 to 8
KH15-K-
□□
terminal box
(large)
G1/2
15a
φ9
φ8 to 9
15b
(standard)
φ10
φ9 to 10
15c
φ11
φ10 to 11
2) Process the end of the cable in accordance with Figure 8-1 and 8-2 below for each model number.
< KH15-J
□□
(small terminal box) >
Strip 60 mm of the cable sheath and 7 to 9 mm of the core coat. Provide shielding at the bottom
end of the sheath. Cable that fits the terminal block is AWG24 to 12 (0.2 to 3.3 mm
2
).
< KH15-K
□□
(large terminal box) >
Strip 80 mm of the cable sheath, and then strip the core coat to fit the length of the M3 crimp
terminal to be used. Mount the crimp terminal to the wiring terminal. Provide shielding at the
bottom end of the sheath.
< Table 8-1 >
< Figure 8-2 >
80
7 or
l
es
s
For M3
< Figure 8-1 >
(60)
7 to 9