EMR-9 Quick Start Guide
__________________________________________________________________________________________________________________________
18
(00175)D
2) Point out a position where the CAL mark
appears on the view image using a laser pointer
or a something stick, such as a pen.
Tips:
9
It is easier to calibrate on a wall or a fixed
plane.
9
If a laser point may not be visible on the
display, use the stick to point out the CAL
mark.
3) When the subject see the pointed
CAL mark (calibration target point),
push the CUE
key (or the Cue
switch). Then the calibration target
point moves to next CAL mark.
4) Repeat seeing the pointed CAL mark and
pushing the CUE
key until end of calibration
(9 calibration points).
Upon completion of the calibration, an eye mark
appears on the view image.
Remarks:
If the subject fails to see the CAL mark, push the
FUNCTION
key to go back the previous
CAL mark.
If the CAL mark doesn’t move even push the
CUE
key, the Pupil and Purkinje images
may not be detected properly. Please cancel the
calibration mode (push the CAL
key again)
and readjust binary image level as per 2.5).