NB-603WS  is  HEPA filtered clean bench  and  UV  PCR  Workstation. This  work station  is  bench-top  type, 
made of metal framework, polycarbonate walls and working surface made of stainless steel. With open UV 
lamps and horizontal air circulation through HEPA filter, this bench is used during operation with DNA/RNA 
sample especially, for PCR work using Thermal Cycler.   

And  also,  this  is  used  for  the  research  which  require  general  clean  area  but,  not  proper  for  sample 
handling which require Biological Safety.       

 

 

CLASS 100 Environment with HEPA filter.   

 

Vertical Laminar Air flow without pre-filter and re-circulation 

 

Polycarbonate side walls and full open door with foldable hand access 
(17cm from bottom) 

 

Long Life Time UV Lamp (Average 4,000 hours) 

 

Ozone free high-density UV decontamination 

 

Stainless Steel Work Surface 

 

Low Noise Blower Motor (Cross flow type Fan) 

 

 
 

 

ITEM

 

NB-603WS

 

Exterior Dimensions

 

700(W) X 650(D) X 940(H)mm 

Working Area Dimensions

 

675(W) X 570(D) X 650(H)mm 

Illumination

 

F.L 36W X 1ea 

Sterilization

 

UV 15W X 1ea 

Blow Fan Type

 

Cross Flow Fan 

Main Filter

 

HEPA Filter (0.3

  99.99%) 

Exhaust Filter

 

None 

Flow Type / Rate

 

Vertical Laminar Flow (ONLY) 
/ Exhaust (Without Filtering) 

Purity

 

Class 100 

Door Type

 

Lift-Up Open 

Material

 

 

Inner 

Stainless steel (SUS304) 

Outer 

Clear Polycarbonate 

Power

 

110/220V, 50/60Hz, 100W 

Weight

 

66kg 

FEATURES 

 

SPECIFICATION 

   

 

Содержание NB-603WS

Страница 1: ...Clean Bench PCR Workstation NB 603WS OPERATING MANUAL...

Страница 2: ...once the unit is proved to be defective by wrong workmanship after NBIOTEK or reliable distributor s examination 2 The warranty period is 1 year from date of installation or 1 and Half year from the d...

Страница 3: ...of the product Do not make the power plug be pressed by back of the product A space between the product and the plug must be 20cm at least The power outlet must be only for this product Using various...

Страница 4: ...and do not install at a hot place or a place that is near an electric heat The The proper room temperature is 20 30 Humidity 40 60 RH Do not put inflammable substances near the product especially usi...

Страница 5: ...oor and body Do not wash the product with excessive quantity of water thinner benzene and Petroleum It can cause an electric leakage and malfunction or damage of the surface Avoid direct UV light to y...

Страница 6: ...ithout pre filter and re circulation Polycarbonate side walls and full open door with foldable hand access 17cm from bottom Long Life Time UV Lamp Average 4 000 hours Ozone free high density UV decont...

Страница 7: ...To open exhaust holes please turn both screws to the right and slide them to the right side To close exhaust holes please slide screw to left side and turn both screws to the left CONFIGURATION...

Страница 8: ...les HEPA Filter The top of the clean bench is equipped with the HEPA filter efficient up to 99 99 at 0 3 Before use of work station please make chamber clean with soft fabric and ethanol And then cond...

Страница 9: ...Don t expose your body skin eyes to UV light UV light is only controlled manually So ensure to turn off UV light when opening door CONTROL PANEL...

Страница 10: ...60Hz Operation 1 Turn on the power s w to operate the product 2 Turn on the UV lamp about 30 minutes before the experiment for sterilization of the chamber 3 Turn off the UV lamp to start the experim...

Страница 11: ...l If necessary wrap it and store it This helps minimize contamination on HEPA Filter Note 1 HEPA FILTER Replacement time for HEPA filter differs depending on time of use and clean condition of room wh...

Страница 12: ...Corrective Action The unit does not operate Make sure the power cord is plugged Make sure the switch is on Check the power outage Decrease of clean air inflow Check the contamination of the filters t...

Отзывы: