Clean Bench (PCR Workstation)
NB-603WS
OPERATING MANUAL
Страница 1: ...Clean Bench PCR Workstation NB 603WS OPERATING MANUAL...
Страница 2: ...once the unit is proved to be defective by wrong workmanship after NBIOTEK or reliable distributor s examination 2 The warranty period is 1 year from date of installation or 1 and Half year from the d...
Страница 3: ...of the product Do not make the power plug be pressed by back of the product A space between the product and the plug must be 20cm at least The power outlet must be only for this product Using various...
Страница 4: ...and do not install at a hot place or a place that is near an electric heat The The proper room temperature is 20 30 Humidity 40 60 RH Do not put inflammable substances near the product especially usi...
Страница 5: ...oor and body Do not wash the product with excessive quantity of water thinner benzene and Petroleum It can cause an electric leakage and malfunction or damage of the surface Avoid direct UV light to y...
Страница 6: ...ithout pre filter and re circulation Polycarbonate side walls and full open door with foldable hand access 17cm from bottom Long Life Time UV Lamp Average 4 000 hours Ozone free high density UV decont...
Страница 7: ...To open exhaust holes please turn both screws to the right and slide them to the right side To close exhaust holes please slide screw to left side and turn both screws to the left CONFIGURATION...
Страница 8: ...les HEPA Filter The top of the clean bench is equipped with the HEPA filter efficient up to 99 99 at 0 3 Before use of work station please make chamber clean with soft fabric and ethanol And then cond...
Страница 9: ...Don t expose your body skin eyes to UV light UV light is only controlled manually So ensure to turn off UV light when opening door CONTROL PANEL...
Страница 10: ...60Hz Operation 1 Turn on the power s w to operate the product 2 Turn on the UV lamp about 30 minutes before the experiment for sterilization of the chamber 3 Turn off the UV lamp to start the experim...
Страница 11: ...l If necessary wrap it and store it This helps minimize contamination on HEPA Filter Note 1 HEPA FILTER Replacement time for HEPA filter differs depending on time of use and clean condition of room wh...
Страница 12: ...Corrective Action The unit does not operate Make sure the power cord is plugged Make sure the switch is on Check the power outage Decrease of clean air inflow Check the contamination of the filters t...