Page 16 / 17
No Function! Just skip this stage by pressing set button once.
b. Fourth
Click “CAL/SET”: CO2 Density Calibration
Channel 4 is at CO2 density calibration stage.
Press UP (▲) as much as difference from measured value by precise analyzer if it is higher.
Press DOWN (▼) as much as difference from measured value by precise analyzer if it is lower.
Ex) If measured CO2 value is 5% and Display shows 4%, then press up 1%.
c. Fifth Calibration Setting: Heating Control
NOTE:
Channel 5 is to set heating control point and is designed to prevent a significant temperature
overshoot. This mode is preprogrammed in the factory and should NOT be adjusted by the user.
d. Sixth Calibration Setting: CO2 Gas Supply Control
NOTE:
Channel 6 is to set the values associated with the CO
₂
control solenoid.
This mode is preprogrammed in the factory and should NOT be adjusted by the user.
* When above stages are cleared, please press CAL/SET button to save all new values.