Myron L
®
Company –
ULTRAPEN™ PT5
– Quick Start Guide
Dissolved Oxygen (DO) and Temperature Pen
PT5QSI 08-17
2
V. AIR CALIBRATION (continued)
6.
Push and hold the push button to enter menu mode.
7.
Release the push button while “CAL” is displayed. The display will
alternate between “CAL” and “PUSHnHLD”.
8.
Push and hold the button then release while “Air CAL” is displayed.
9.
The PT5 will perform the calibration. When calibration is complete, the
display will indicate “SAVEd” then turn OFF.
10.
To verify calibration follow steps 1 thru 5 above (taking a
measurement) at 25 ºC, the readings should be 104% ±2%.
NOTE:
The PT5
MUST
be recalibrated any time: the DO sensor is
replaced; the DO membrane or the electrolyte is replaced, or; the Zinc
electrode is cleaned
VI. DO Measurement
CAUTION
: For best results when measuring solution at the extremes of
the specified temperature or DO range
ALWAYS
perform a
TEMPERATURE CONTROLLED, FULL CALIBRATION as described in
the full PT5 Operation Manual. Allow the PT5 to equilibrate by
submerging the sensor in the sample solution for 1 minute prior to taking
a measurement.
•
ALWAYS make sure that the DO Sensor is completely submerged in
the solution being tested.
•
If the expected measurement is < 3 ppm it is particularly important to
perform the ZERO calibration described in the full PT5 Operation
Manual.
1.
Push and release the push button to turn the PT5 ON.
2.
Dip PT5 sensor in solution to be tested and swirl.
3.
Continue to swirl the PT5 in solution while measurement is taking
place.
The following table explains what the LED Indicator light signals indicate
and gives the duration of each signal.
NOTE
: You can turn OFF the unit at any time during live measurement
with a simple press and release of the push button.
LED Signal
Action
Duration
Rapid Flashing
Dip pen in solution and swirl.
6 sec
Slow Flashing
Measurement in process; continue
to swirl.
In
HOLD
mode real-time readings
are displayed until the LED is ON
solid.
In
LIVE
mode real-time readings
are displayed until the PT5 turns
OFF.
20 -120 sec in
Hold mode
Up to 5 min in
LIVE mode
Solid Light
(Hold mode only
)
Measurement is complete. Values
are displayed until the PT5 turns
OFF.
6 sec
MAINTENANCE
I. Routine Maintenance
1.
After each use
ALWAYS
rinse the DO sensor with clean water (preferably DI, RO, or
distilled water), then carefully blot the sensor membrane with a soft, clean, lint-free
cloth to remove any water drops.
2. ALWAYS
replace the hydration cap on the DO sensor after each use.
3.
To replace the membrane cap and/or refill electrolyte solution follow the SENSOR
SETUP instructions.
4.
Do not drop, throw or otherwise strike the PT5. This voids the warranty.
5.
Do not store the PT5 in a location where the ambient temperatures exceed its
specified Operating/Storage Temperature limits.
NOTE:
For instructions on cleaning the DO sensor, performing a sensor check,
troubleshooting, or storing the PT5, download the full PT5 Operation Manual from
www.myronl.com.
II. Battery Replacement
1.
In a
CLEAN DRY
place unscrew the battery cap in a counter-clockwise direction.
2.
Slide the cap and battery housing out of the PT5.
3.
Remove the depleted battery from its housing.
4.
Insert a new battery into the battery housing oriented with the negative end touching the spring.
5.
Align the groove along the battery housing with the guide bump inside the PT5 case and slide the battery housing back in.
6.
Screw the battery cap back on in a clockwise direction. Do not over tighten.
SPECIFICATIONS
Measurement Ranges
DO Concentration: 0-20 ppm, DO Saturation 0-200%
Temperature: 0-71°C / 32-160°F
Accuracy
1
Concentration:
±0 .2ppm or ± 2% of the reading, whichever is greater.
Saturation:
0.0% to
≤
100.0%: ± 20 counts;
>100%: ± 2% of the reading.
Temperature:
± 0.1°C
Time to Reading
Stabilization
10 - 180 seconds
Operating / Storage
Temp
0-55ºC / 32-131ºF
Resolution
DO Concentration: 0.01 ppm, DO Saturation 0.1%
Temperature 0.1ºC/F
Altitude Adjustment
-2,000 ft. to 15,000 ft. in increments of 100 ft.
Salinity Adjustment
0 - 50 ppt in increments of 1 ppt
Power Consumption
Active Mode: 33 - 140 mA, Sleep Mode: 2 µA
Battery Type
N type, Alkaline, 1.5 V
Temp Compensation
Automatic Temperature Compensation for Membrane
permeability
Calibration Methods
1 or 2 Point; 0 ppm DO Calibration Solution; Air or Water
Physical Dimensions /
Weight
17.15 cm L x 1.59 cm D or 6.75 in. L x .625 in. D;
Weight: 54g or 1.94oz
Case Material:
Anodized Aircraft Aluminum with Protective Coating
Water Resistance
IP67 and NEMA 6
EN61236-1: 2006 -Annex A: 2008; 1999/5/EC Annex III: Electrostatic discharge to
case of instrument may cause PT5 to spontaneously power ON. In this case, the
PT5 will power OFF after several seconds.
1
Accuracy after Temperature Controlled, Full Calibration. For this procedure and for more detailed accuracy specs download the full PT5 Operation Manual from www.myronl.com.
MYRON L
®
COMPANY
2450 Impala Drive
!
Carlsbad, CA 92010-7226
!
Phone: +1-760-438-2021
E-Mail: Customer Service – [email protected]
!
Technical Support – [email protected]
Website: www.myronl.com