ULTRAPEN™ PT3 – Quick Start Guide 

ORP and Temperature Pen

 

PT3QSI  07-21  

 III. CALIBRATION: 

 

Myron L

®

 Company recommends performing a calibration of the PT3 twice a month or any time measurements are not as 

expected.

    

  The PT3 has embedded automatic temperature compensation in calibration mode (from 15ºC to 30ºC) for three calibration solutions (80 mV and 

260 mV Quinhydrone as well as 470 mV Light’s Solution).  
For more information, visit www.myronl.com/pt3.htm. 

  Small bubbles trapped in the sensor may give a false calibration.   

 

1. 

Rinse the PT3 sensor in ORP solution.  

  The PT3 will automatically identify which ORP standard solution you are using. 

2. 

Press and release the push button to turn the PT3 ON. 

3. 

Push and Hold the button down.   

4. 

Release the button when “CAL” is displayed. 

5. 

Grasp the pen near the battery cap to avoid sample contamination. 

6.

  While the LED flashes rapidly, dip the pen in fresh sample solution so that the sensor is completely submerged.  

  If you do not submerge the sensor in solution before the flashing slows, allow the pen to power OFF and retake the reading. 

7.

  While the LED flashes slowly, swirl the pen around to remove bubbles, keeping the sensor submerged.  

  If possible, keep the pen at least 1 inch (2½ cm) away from sides/bottom of container. 

8.  

When the LED turns ON solid, remove the pen from solution.  

  The display will read “CAL SAVED” and the PT3 will turn OFF. 

  If “Error” appears on the display, check to make sure you are using a proper, Myron L

®

 Company, ORP Standard Solution. 

  If the solution is correct, clean the sensor, as described in the full PT3 Operation Manual downloadable from the Myron L

®

 Company website. 

FACTORY CALIBRATION

Factory Calibration (FAC CAL) resets the PT3 to its factory settings.  For instructions on how to perform a FAC CAL 

download the full PT3 Operation Manual from the Myron L

®

 Company website. 

 

 

MAINTENANCE 

I. BATTERY REPLACEMENT: 

  

1. 

In a clean, dry place unscrew the battery cap in a counter-clockwise motion.  

2. 

Slide the cap and battery housing out of the pen.  

3. 

Remove the depleted battery from its housing.  

4.  I

nsert a new battery into the battery housing oriented with the negative end 

touching the spring.  

5. 

Align the groove along the battery housing with the guide bump inside the 
pen case and slide the battery housing in.  

6. 

Screw the pen cap back on in a clockwise direction.  Do not over tighten.  

II. ROUTINE MAINTENANCE  

1.

  After each use, 

ALWAYS

 rinse the sensor with clean water (preferably DI, 

RO, or purified water), then carefully blot the sensor with a soft,  clean cloth 
to remove any water drops.  

2.

 

ALWAYS

 replace the Soaker Cap (filled half way with Myron L

®

 Company 

Storage Solution) on the sensor after each use. 

DO NOT 

push the cap past the Cap Stop.  

3.

  Do not drop, throw or otherwise strike the PT3. This voids the warranty.  

4.

  Do not store the PT3 in a location where the ambient temperatures exceed its specified Operating/Storage Temperature limits.  

II. SENSOR CLEANING AND REPLACEMENT:

  

For sensor cleaning and conditioning instructions download the full PT3 Operation Manual from 

the Myron L

®

 Company website.

 

 

 

SPECIFICATIONS 

Measurement Ranges

 

ORP-1000mV – 1000mV,  Temp: 0°-71° C / 32-160° F 

ORP Accuracy 

 

± 10mV

 

Resolution

 

ORP 1mV,    Temp: ± 0.1ºC / ± 0.1ºF

 

Temp. Accuracy 

± 0.1ºC / ± 0.1ºF

 

Operating/Storage 
Temp

 

0º - 55ºC / 32º - 131ºF

 

Calibration Solutions 

80mV, 260mV, 470mV 

Temp Compensation

 

Automatic In Calibration Mode From 15ºC to 30ºC

 

Time to Stabile 
Reading  

10 - 45 seconds

 

Power Consumption

 

Active Mode: 37 mA, Sleep Mode: 2 

μ

A

 

Battery Type

 

N type, Alkaline 1.5 V

 

Physical Dimensions  

17.15 cm L x 1.59 cm D or 6.75 in. L x .625 in. D;   
Weight: 50.4g / 1.78oz. 

Case:  

Anodized Aircraft Aluminum with Protective Coating 

Water Resistance

 

IP67 and NEMA 6

 

EN61236-1: 2006 - Annex A: 2008; Electrostatic discharge to case of  
instrument may cause PT3 to spontaneously power ON. In this case,  
the PT3 will power OFF after several seconds. 

 

MYRON L

®

 COMPANY

 

2450 Impala Drive   

u

 

Carlsbad, CA 92010-7226   

u

 

Phone: +1-760-438-2021 

E-Mail: Customer Service – [email protected]     

u

 

Technical Support – [email protected] 

Website: www.myronl.com  

 

 

Отзывы: