 9P Quick Start Guide 

 

Handheld Conductivity, Resistivity, TDS, pH, ORP, Hardness, Alkalinity, LSI, Free Chlorine Equivalent (FC

E™

), and Temperature Meter

 

MYRON L

®

 COMPANY

  

2450 Impala Drive   

!

 

Carlsbad, CA 92010-7226   

!

 

Phone: +1-760-438-2021 

E-Mail: Customer Service – [email protected]     

!

 

Technical Support – [email protected] 

 

Website: www.myronl.com

  

9PQSI 06-19

 

NOTES:  

  When finished making Conductivity, Resistivity, TDS, Alkalinity or Hardness 

measurements, rinse the cell cup with clean water (preferably DI, RO, or 
distilled). 

  When finished measuring pH, ORP or FC

E

, rinse sensor well with clean water 

(preferably DI, RO, or Distilled), refill it with Myron L

®

 pH/ORP Sensor Storage 

Solution then reinstall the protective cap. 

II. Solution Mode Selection 

1.  Press         ,          or        , to select which parameter is having its solution  

type changed. 

 

2.  Press and hold         about 3 seconds or until "SEL" appears on the display. 

3.   Use the         or         key to select type of solution desired. The possible  

 

solution types are KCI, NaCl or 442 Natural Water™ Standard. 

4.  Press         to accept new solution type. 

NOTE: The solution mode selection determines the salt solution characteristic 
used to model temperature compensation for conductivity, resistivity and TDS. 

 

CALIBRATION 

CALIBRATION SOLUTIONS 

Mode 

Standard Solution or Buffer 

Conductivity KCl 

KCl: 7000 

µ

TDS 442™ 

442™: 3000 PPM 

pH 

4.0, 7.0 and 10.0 buffer 

Alkalinity 

100 ppm Alkalinity Standard Solution  

Hardness 

200 ppm Hardness Standard Solution  

FC

E™

 ORP calibration is electronic and based on pH Cal Results 

Resistivity is the reciprocal of Conductivity. To calibrate resistivity, calibrate conductivity 
for the solution type you wish to measure

 

I. CONDUCTIVITY OR TDS CALIBRATION 

1.  Rinse cell cup 3 times with proper standard solution. 
2.  Refill cell cup with fresh, standard solution. 

3.  Press         or        , then press         , "CAL" will appear. 

4.  Press         or         until display agrees with standard solution. 

5.  Press         to accept value. 

NOTE: A deviation of more than 10% from standard solution value will  
produce a "FAC" on the display. Press         to accept factory calibration, 
or clean the cell and recalibrate. 

II. ALKALINITY and HARDNESS CALIBRATION  

  For complete instructions on performing these calibrations, download the full 

Ultrameter III 9P Operation Manual.  

III. pH CALIBRATION 

1.  Rinse sensor well and cell cup 3 times with 7.0 buffer solution. 
2.  Refill sensor well and cell cup with 7.0 buffer solution. 
3.  Press        , then        , "CAL", "7" and "BUFFER" will appear on the display. 

4.  Press         or         until display agrees with buffer value. 

5.  Press          once to accept the value. 

6.  Rinse 3 times with pH 4.0 or pH 10.0 buffer solution. 
7.  Refill and repeat steps 4 & 5. 
8.  Rinse 3 times with opposite buffer solution, refill and repeat steps 4 & 5 for a  

3-point calibration, or press         to exit. 

9.  When calibration is complete, rinse the sensor well then fill it with Myron L

®

 

pH/ORP Sensor Storage Solution and reinstall the protective cap.  

IV. ORP/ FC

E

 CALIBRATION 

  ORP electrodes rarely give false readings without problems in the reference 

electrode.  

  For this reason, and because calibration solutions for ORP are highly reactive 

and potentially hazardous, your Ultrameter III has an electronic ORP 
calibration.  

  This causes the zero point on the reference electrode to be set whenever pH 7 

calibration is done.

 

MAINTENANCE

I. BATTERY REPLACEMENT (LOBATT) 

1. 

Clean and dry the instrument THOROUGHLY.  

WARNING:

 

ONLY

 open the 9P in a clean and dry environment! 

2. 

Remove the four (4) bottom screws. 

3. 

Open the instrument CAREFULLY.  

4. 

Carefully detach the battery from the circuit board.  

5. 

Replace with a new 9-volt alkaline battery.  

6. 

Replace bottom, ensuring the sealing gasket is installed in the groove of the 
top half of case. 

7.

  Reinstall the 4 screws, tightening them evenly and securely. 

II. CAUTIONS 

  Solutions in excess of 71°C/160°F should not be placed in the cell cup area; 

this may cause damage. 

  The pH/ORP sensor may fracture if the 9P temperature is allowed to go below 

0°C/32°F. 

  The Conductivity cell cup should be kept clean. Flushing with clean water 

following use will prevent buildup on electrodes. If very dirty samples  
(particularly scaling types) are allowed to dry in the cell cup, a film will form 
reducing accuracy.  

  The pH/ORP sensor in your 9P

 

should not be allowed to dry out.  

  Whenever the pH/ORP sensor is not in use make sure it is filled with  

Myron L

®

 Company pH/ORP Sensor Storage Solution and that the     

protective cap is in place. 

III. CLEANING SENSORS 

A.  Conductivity, TDS, Resistivity, Alkalinity, Hardness, and LSI 

  When there are visible films of oil, dirt, or scale in the cell cup or on the 

electrodes, or if readings are not as expected, clean cell using isopropyl 
alcohol or a foaming non-abrasive household cleaner. 

  Rinse out the cleaner, and your 9P is ready for accurate calibration then 

measurements. 

B.  pH/ORP

:  If the pH/ORP sensor dries out or becomes dirty, it may be 

cleaned and/or reconditioned.  For instructions on performing these 
operations, download the full 9P Operation Manual from the 
Myron L

®

 Company website. 

IV. pH/ORP Sensor Replacement 

Order model RPR. Be sure to include the model and serial number of your 
instrument to ensure receipt of the proper type.  

NOTE

: Complete installation 

instructions are provided with each replacement sensor.

SPECIFICATIONS 

Spec 

CONDUCTIVITY 

TDS  

RESISTIVITY 

pH 

ORP 

ALKALINITY 

HARDNESS 

LSI 

FC

E™

 

TEMP 

Ranges 

0 - 9999 

µ

S/cm  

10 - 200 mS/cm 

In 5 autoranges

 

0 - 9999 ppm 

10 - 200 ppt  

In 5 autoranges

 

10 k

Ω

 - 30 M

Ω

 

0-14 pH 

±999 mV 

10-800 ppm 

0-1710 ppm 

(0-100 grains) 

-10 to +10 

0.00 – 9.99 ppm 

ORP = 350mV - 725 mV, <9.9 pH 

ORP = 725 mV - 825 mV, <8.9 pH 

0°C – 71°C 

32°F – 160°F 

Resolution 

0.01 (<100 

µ

S) 

0.1  (<1000 

µ

S) 

1 (<10 mS) 

0.01 (<100 mS) 

0.1 (<200 mS) 

0.01 (<100 ppm) 
0.1 (<1000 ppm) 

1 (<10 ppt) 

0.01 (<100 ppt) 

0.1 (<200 ppt) 

0.01 (<100 k

Ω

0.1 (<1000 k

Ω

0.1 (>1 M

Ω

0.01 pH

 

1 mV 

0.1 ppm 

0.1 (<1000 ppm) 

1 (<1710 ppm) 

0.1 (<100 grains) 

0.1 

0.01 ppm 

0.1 °C / °F 

Accuracy 

±

1% of reading 

±

1% of reading 

±

1% of reading 

± 0.01 pH 

±1 mV 

 

 

 

±0.3 ppm <1.0 ppm 
±0.2 ppm 

1.0 ppm 

±0.1 °C 

COND/TDS Ratios  Programmed: KCl, NaCl or 442™, Adjustable 0.20 -07.99 

 

 

 

 

 

 

 

 

Temperature 

Compensation 

Auto: 0-71°C, 32-160°F, Adjustable from 0 – 9.99% / °C 

Auto: 0-71°C, 

32-160°F 

 

 

 

 

 

 

 